This work presents an analysis of a total hip prosthesis that failed in service prematurely within 5 years. The prosthesis presented a fractured stem that was extracted from a 46 year old male patient, 75 kg weight and 1.76 m height. In order to determine the origin of the failure, the femoral stem component was analyzed by means of visual inspection; optical microscope (OM), scanning electronic microscopy (SEM) and energy disperse spectroscopy (EDS). The investigation showed that the crack had originated due to a high stress concentration on the anterolateral corner section of the stem. Any discontinuity or defect on the prosthesis surface in this location acted as preferential site to nucleate a crack which propagated by fatigue until the cross section was not able to sustain an eventual high load produced for the active patient.
机译:这项工作提出了对总髋关节假体的分析,在5年内过早发生服务。 假体呈现出一种骨折茎,从46岁的男性患者,75千克重量和1.76米高。 为了确定失败的起源,通过目视检查分析股骨干组分; 光学显微镜(OM),扫描电子显微镜(SEM)和能量分散光谱(EDS)。 研究表明,由于茎的前外角部分上的高应力浓度,裂缝起源于裂缝。 在该位置的假体表面上的任何不连续性或缺陷作用为核心的优先部位,以核心通过疲劳繁殖,直到横截面不能维持为活性患者产生的最终的高负荷。