The generation of polyethylene wear debris, and the subsequent tissue reaction to such debris is considered to be a limitation in the long-term survival of shoulder arthroplasties. The purpose of this study was to investigate, for the first time, the wear of a novel PyroCarbon-on-Polyethylene (PyCoP) shoulder arthroplasty system. A 5 million cycle wear test was performed on PyroCarbon humeral heads, which were articulated against commercially available polyethylene glenoid insert components to form an anatomic total shoulder arthroplasty (aTSA). A "Repeat-motion-load" physiological combined cycle was applied using the unique Newcastle Shoulder Wear Simulator. Wear was assessed gravimetrically, and the change of the surface roughness was measured with a non-contacting profilometer. The mean wear rate of the ultra-high molecular weight polyethylene (UHMWPE) components was 19.3 +/- 9.5 mm(3)/million cycles after 5 million cycles of testing. The roughness value, S-a, of the UHMWPE glenoid inserts, reduced, changing from 296 +/- 28 nm S-a to 32 +/- 8 nm S-a. In contrast, the mean roughness of the PyroCarbon humeral heads remained in the same range (21 +/- 2 nm S-a to 20 +/- 10 nm S-a). There was no reduction in weight (no measurable wear) of the PyroCarbon humeral heads over the duration of testing. This study is the first to describe the wear performance of UHMWPE glenoid inserts against PyroCarbon humeral heads. No significant difference in the wear of UHMWPE was found in comparison with published studies.
机译:聚乙烯磨损碎片的产生以及随后对这种碎片的组织反应被认为是肩关节塑化术的长期存活中的限制。本研究的目的是首次进行研究,这是一种新型丙酮酰基聚乙烯(PYCOP)肩关节成形术系统的磨损。对Pyrocarbon肱骨头进行了500万个循环磨损试验,其针对市售的聚乙烯肾盂骨插入部件铰接,形成一种解剖总肩部轴承术(ATSA)。使用独特的纽卡斯尔肩部磨损模拟器应用“重复运动负荷”生理组合循环。热分析评估磨损,用非接触式轮廓仪测量表面粗糙度的变化。超高分子量聚乙烯(UHMWPE)组分的平均磨损率为19.3 +/- 9.5mm(3)/百万次循环,经过500万次的测试后。 UHMWPE胶质套管插入件的粗糙度值S-A降低,从296 +/- 28nm S-a变为32 +/- 8nm s-a。相反,Pyrocarbon肱骨头的平均粗糙度保持在相同的范围(21 +/- 2nm S-A至20 +/- 10nm S-A)。在测试期间没有减轻Pyrocarbob肱骨头的重量(无可测量磨损)。本研究是第一个描述UHMWPE胶质螺纹插入锁定Pyrocarbon肱骨头的磨损性能的方法。与公布的研究相比,发现了UHMWPE的磨损没有显着差异。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号