New pocket echocardiography device is interchangeable with high-end portable system when performed by experienced examiners.

机译:由经验丰富的检查员执行时,新型袖珍超声心动图设备可与高端便携式系统互换。

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

BACKGROUND: Cardiovascular status is a crucial determinant in the pre-operative assessment of patients for surgery as well as for the handling of patients with acute illness. We hypothesized that focus-assessed transthoracic echocardiography (FATE) could be performed with the subject in the semi-recumbent position. The aim was also to test whether the image quality of Vscan is interchangeable with a conventional high-quality portable echocardiography system. Furthermore, we evaluated the time needed to achieve an interpretable four-chamber view and to complete a full FATE examination. METHODS: Sixty-one subjects were included. All subjects were examined in accordance with the FATE protocol in the semi-recumbent position on two different systems: the novel Vscan pocket device and the high-quality portable Vivid i system. Two evaluations were performed. In group A (n=30), the focus was on image quality. In group B (n=31), the focus was on the time consumed. RESULTS: Group A: All patients (100%) had at least one image suitable for interpretation and no significant difference in image quality (P=0.32) was found between the two different systems. Group B: The mean value for the total time consumed for a full FATE was 69.3 s (59.8-78.8) on the Vscan and 63.7s (56.7-70.8) on the Vivid i, with no significant difference among the scanners (P=0.08). CONCLUSION: The Vscan displays image quality interchangeable with larger and more expensive systems. The apparatus is well suited for performing a FATE examination in a 1-day surgery setting and could very well also be applicable in almost any situation involving patients with acute illness.
机译:背景:心血管状况是手术患者术前评估以及急性疾病患者处理的关键决定因素。我们假设可以对受试者处于半卧位进行聚焦评估的经胸超声心动图(FATE)。目的还在于测试Vscan的图像质量是否可以与传统的高质量便携式超声心动图系统互换。此外,我们评估了实现可理解的四腔视图并完成完整的FATE检查所需的时间。方法:纳入61名受试者。所有受试者均按照FATE协议在两个不同系统的半卧姿位置进行了检查:新颖的Vscan袖珍设备和高质量的便携式Vivid i系统。进行了两次评估。在A组(n = 30)中,重点是图像质量。在B组(n = 31)中,重点是消耗的时间。结果:A组:所有患者(100%)都有至少一张适合于解释的图像,并且在两个不同的系统之间未发现图像质量的显着差异(P = 0.32)。 B组:完整FATE消耗的总时间平均值在Vscan上为69.3 s(59.8-78.8),在Vivid i上为63.7s(56.7-70.8),扫描仪之间无显着差异(P = 0.08) )。结论:Vscan显示的图像质量可以与更大,更昂贵的系统互换。该设备非常适合在1天的手术环境中执行FATE检查,并且也非常适用于几乎所有涉及急性疾病的患者。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号