首页> 外文期刊>Journal of neurointerventional surgery >Use of the 0.027-inch VIA microcatheter for delivery of Pipeline Flex: a technical note
【24h】

Use of the 0.027-inch VIA microcatheter for delivery of Pipeline Flex: a technical note

机译:使用0.027英寸通过微直接管进行管道Flex:技术说明

获取原文
获取原文并翻译 | 示例
           

摘要

Background The Pipeline Embolization Device (PED; Medtronic Neurovascular, Irvine, California, USA) is designed for delivery through a 0.02700 microcatheter. Challenges with the second-generation PED Flex include limited support from the Marksman microcatheter for consistent resheathing and transmission of push forces for device delivery. The VIA27 (Sequent Medical/MicroVention Terumo, Tustin, California, USA) is an alternative 0.02700 microcatheter originally designed for intrasaccular flow diverter delivery. Here we describe our experience with the VIA27 in the delivery of PED Flex.
机译:背景技术管道栓塞装置(PED; Medtronic Neurovancular,Irvine,California,USA)设计用于通过0.02700微导管的输送。 第二代PED Flex的挑战包括来自Marksman Microcatheter的有限支持,以便为装置输送的支撑力的一致性和传输。 VIA27(Sequent Medical / MicroVention Terumo,Tustin,California,USA)是最初设计用于跨置换蓄水器交付的0.02700微导管。 在这里,我们在PED Flex的交付中描述了我们对VIV27的经验。

著录项

  • 来源
  • 作者单位

    Univ Calif Irvine Dept Neurosurg Sch Med UC Irvine Med Ctr 200 S Manchester Ave Suite 210;

    Johns Hopkins Univ Sch Med Dept Neurosurg Johns Hopkins Hosp Baltimore MD USA;

    Johns Hopkins Univ Sch Med Dept Neurosurg Johns Hopkins Hosp Baltimore MD USA;

    Johns Hopkins Univ Sch Med Dept Neurosurg Johns Hopkins Hosp Baltimore MD USA;

    Johns Hopkins Univ Sch Med Dept Neurosurg Johns Hopkins Hosp Baltimore MD USA;

    Johns Hopkins Univ Sch Med Dept Neurosurg Johns Hopkins Hosp Baltimore MD USA;

    Johns Hopkins Univ Sch Med Dept Neurosurg Johns Hopkins Hosp Baltimore MD USA;

  • 收录信息
  • 原文格式 PDF
  • 正文语种 eng
  • 中图分类 神经病学与精神病学;
  • 关键词

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号