Newly-Developed Positron Emission Mammography (PEM) Device for the Detection of Small Breast Cancer

机译:用于检测小乳腺癌的新开发的正电子发射乳房X线照片(PEM)装置

获取原文
获取原文并翻译 | 示例
           

摘要

Positron emission mammography (PEM) has higher detection sensitivity for breast cancer compared with whole-body positron emission tomography (PET) due to higher spatial resolution. We have developed a new PEM device with high resolution over a wide field of view. This PEM device comprises novel scintillation crystals, praseodymium-doped lutetium aluminum garnet (Pr:LuAG). In the present study, the clinical use of the newly developed PEM for the detection of small breast cancer was compared with that of the conventional PET-computed tomography (PET/CT). Eighty-two patients with breast cancer less than 20 mm (UICC T1) participated in this study, including 23 patients with T1a or T1b breast cancer (less than 10 mm). Histologically-proved lesions were examined by PET/CT and PEM on the same day after injection of [F-18]fluoro-2-deoxy-2-fluoro-D-glucose ([F-18]FDG), a marker of glycolytic activity. The newly developed PEM showed better sensitivity of cancer detection compared with PET/CT especially in case of the small T1a or T1b lesions. Moreover, when the conventional PET/CT and new PEM were combined, the detection sensitivity with [F-18]FDG molecular imaging for T1 (N = 82) and T1a plus T1b breast cancer (N = 23) were 90% and 70%, respectively. The uptake of [F-18]FDG was proportional to the histological malignancy of breast cancer. Using the newly-developed PEM with [F-18]FDG, we are able to identify and characterize exactly the small breast tumors less than 10 mm in combination with the conventional PET/CT. These data indicate that PEM and PET/CT are synergic and complementary for the detection of small breast cancer.
机译:由于空间分辨率较高,正电子癌乳腺癌(PEM)对乳腺癌的检测敏感性较高。我们开发了一种具有高分辨率的新PEM器件,在广泛的视野中。该PEM器件包括新型闪烁晶体,镨掺杂铝石榴石(PR:Luag)。在本研究中,将新开发的PEM用于检测小乳腺​​癌的临床应用与传统的PET计算机断层扫描(PET / CT)进行了比较。八十二名乳腺癌患者少于20毫米(UICC T1)参与了本研究,其中包括23例T1A或T1B乳腺癌(小于10毫米)。在注射[F-18]氟-2-脱氧-2-氟-d-氟-d-氟-d-氟-d-氟-d-荧光([F-18] FDG)的同一天,通过PET / CT和PEM检查组织学验证的病变,糖酵母的标记物活动。与PET / CT相比,新开发的PEM表现出更好的癌症检测敏感性,特别是在小T1A或T1B病变的情况下。此外,当组合常规PET / CT和新PEM时,对T1(n = 82)和T1A加T1B乳腺癌(n = 23)的检测灵敏度为T1(n = 82)和T1a加T1b乳腺癌(n = 23)为90%和70% , 分别。 [F-18] FDG的摄取与乳腺癌的组织学恶性成比例。使用具有[F-18] FDG的新开发的PEM,我们能够与传统的PET / CT组合识别和表征小于10mm的小乳腺肿瘤。这些数据表明PEM和PET / CT是用于检测小乳腺​​癌的协同和互补性。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号