MDA/2019/025 IntelliVue MX40 patient-worn monitors - increased power consumption and no visual or audible alarms when batteries are low. Manufactured by Philips -devices may lose power earlier than expected and users may not realise the loss of monitoring due to no alarm, which could contribute to a delay in emergency treatment.MDA/2019/024 Dialog +haemodialysis machines with software versions 9.xx (excluding versions 9.18, 9.1 A and 9.1 B) -software and hardware upgrade required. Manufactured by B. Braun Avitum AG - Malfunction of the temperature sensor can result in temperature of the dialysis fluid to be more than ±7 °C outside the programmed values, which can lead to inadequate treatment.