Validation of a cuff-based device for measuring carotid-femoral pulse wave velocity in children and adolescents

机译:用于测量儿童和青少年颈动脉脉搏波速度的携带基于袖带的装置

获取原文
获取原文并翻译 | 示例
           

摘要

Carotid-femoral pulse wave velocity is associated with arterial stiffness in major elastic arteries, and predicts future cardiovascular events. However, little is known about carotid-femoral pulse wave velocity as a marker of vascular health in children. Semi-automated cuff-based devices for assessing pulse wave velocity are increasingly popular, although these utilize an algorithm developed and validated in adults. Physiological differences between adults and children may, however, reduce the accuracy of cuff-based methods. We sought to determine the accuracy of a cuff-based pulse wave velocity device in healthy children, and determine whether a novel age-appropriate algorithm increases accuracy. We recruited 29 healthy children between the ages of 2 and 20 years. Pulse wave velocity was measured both by using a tonometer on the carotid artery and an inflated cuff on the thigh, and using a tonometer on both the carotid artery and femoral artery as a reference standard. Accuracy of the cuff-based device with its standard algorithm developed in adults, and a novel age-appropriate algorithm corrected for physiological differences in leg pulse wave velocity was assessed with Regression analysis and Bland-Altman plots. Cuff-based device estimates of pulse wave velocity had excellent agreement to the reference standard (Delta = -0.26 ms(-1) [SD 0.44]). The novel age-appropriate algorithm improved the accuracy of the cuff-based method (Delta = 0.02 ms(-1) [SD 0.44]). The cuff-based semi-automatic approach estimates carotid-femoral pulse wave velocity with excellent agreement to the reference standard. However, adjusting the algorithm for known differences in leg pulse wave velocity further improves the accuracy of cuff-based measurement in children and adolescents.
机译:颈动脉 - 股骨脉搏波速度与主要弹性动脉的动脉僵硬有关,并预测未来的心血管事件。然而,关于颈动脉 - 股骨脉搏波速度的少量熟知作为儿童血管健康的标志物。用于评估脉搏波速度的半自动袖带的装置越来越流行,尽管这些算法利用成人开发和验证的算法。然而,成人和儿童之间的生理差异可能降低了基于袖带的方法的准确性。我们试图确定健康儿童中袖带的脉冲波速度装置的准确性,并确定一种新的年龄适当算法是否提高了准确性。我们招募了29岁及20年之间的健康儿童。通过在颈动脉和大腿上使用颈动脉和膨胀的袖口,并在颈动脉和股动脉上使用眼压计作为参考标准,测量脉冲波速度。用成人开发的基于袖带的装置的准确性,并评估了对腿部脉搏波速度的生理差异的新型适当的算法,并用回归分析和Bland-altman图评估了腿部脉冲波速度的生理差异。脉冲的脉冲波速度的估计与参考标准(Delta = -0.26ms(-1)[SD 0.44])具有优异的一致性。新颖的年龄适当的算法提高了基于袖带的方法的准确性(Delta = 0.02ms(-1)[SD 0.44])。基于袖带的半自动方法估计颈动脉 - 股脉冲波速度,具有优异的参考标准。然而,调整腿脉冲波速度已知差异的算法进一步提高了儿童和青少年的基于袖带的测量的准确性。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号