A Semiautomated Method for Measuring the 3-Dimensional Fabric to Renal Artery Distances to Determine Endograft Position After Endovascular Aneurysm Repair

机译:用于测量肾动脉距离的三维织物的半过滤方法,以确定血管内动脉瘤修复后的内血血移植位置

获取原文
获取原文并翻译 | 示例
       

摘要

Purpose: To report a methodology for 3-dimensional (3D) assessment of the stent-graft deployment accuracy after endovascular aneurysm repair (EVAR). Methods: A methodology was developed and validated to calculate the 3D distances between the endograft fabric and the renal arteries over the curve of the aorta. The shortest distance between one of the renal arteries and the fabric (SFD) and the distance from the contralateral renal artery to the fabric (CFD) were determined on the first postoperative computed tomography (CT) scan of 81 elective EVAR patients. The SFDs were subdivided into a target position (0–3 mm distal to the renal artery), high position (partially covering the renal artery), and low position (3 mm distal to the renal artery). Data are reported as the median (interquartile range, IQR). Results: Intra- and interobserver agreements for automatic and manual calculation of the SFD and CFD were excellent (ICC 0.892, p0.001). The median SFD was 1.4 mm (IQR ?0.9, 3.0) and the median CFD was 8.0 mm (IQR 3.9, 14.2). The target position was achieved in 44%, high position in 30%, and low position in 26% of the patients. The median slope of the endograft toward the higher renal artery was 2.5° (IQR ?5.5°, 13.9°). Conclusion: The novel methodology using 3D CT reconstructions enables accurate evaluation of endograft position and slope within the proximal aortic neck. In this series, only 44% of endografts were placed within the target position with regard to the lowermost renal artery. ]]>
机译:目的:报告血管内动脉瘤修复(EVAR)后支架移植队列准确度的三维(3D)评估方法。方法:开发和验证方法,以计算主移植织物与主动脉曲线上的内血症织物和肾动脉之间的3D距离。在81个选修EVAR患者的第一个术后计算断层扫描(CT)扫描上,确定了肾动脉和织物(SFD)之间的最短距离和距离对侧肾动脉的距离和织物(CFD)的距离。将SFD细分为靶位置(远端至肾动脉的0-3mm),高位置(部分覆盖肾动脉),低位置(肾动脉远端,肾脏动脉的肾脏动脉)。数据被报告为中位数(局部范围,IQR)。结果:SFD和CFD的自动和手动计算的内部和Interobserver协议优异(ICC& 0.892,P <0.001)。中位数SFD为1.4毫米(IQR?0.9,3.0),中位数CFD为8.0毫米(IQR 3.9,14.2)。目标位置以44%,高位为30%,低位为26%的患者。朝向更高的肾动脉的内涂层的中值斜率为2.5°(IQR?5.5°,13.9°)。结论:采用3D CT重建的新型方法能够精确评估近端主动脉颈内的内血移植位置和斜率。在该系列中,仅在逆下的肾动脉的目标位置内只有44%的内切移植物。 ]]>

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号