首页> 外文期刊>Journal of digital imaging: the official journal of the Society for Computer Applications in Radiology >Estimates of Patient Radiation Doses in Digital Radiography Using DICOM Information at a Large Teaching Hospital in Oman
【24h】

Estimates of Patient Radiation Doses in Digital Radiography Using DICOM Information at a Large Teaching Hospital in Oman

机译:在阿曼大型教学医院DICOM信息DICOM信息估计患者放射线剂量的估计

获取原文
获取原文并翻译 | 示例

摘要

In this study, we sought to estimate the patient radiation doses in the digital radiography X-ray examinations conducted in a large hospital. The patient exposure factors and kerma-area product (P-KA) were retrospectively recorded via the Digital Imaging and Communications in Medicine (DICOM) header for 547 patients. The entrance surface air kerma (ESAK) was estimated from the measurements of the X-ray tube output and recorded exposure factors, as well as from the console that displayed P-KA as an alternative method. Effective doses were estimated from ESAK and P-KA values using the appropriate conversion coefficient. In the chest PA, chest LAT, cervical spine AP, cervical spine LAT, abdomen AP, pelvis AP, lumbar spine AP, and lumbar spine LAT, the median ESAK (mGy) was found to be 0.13, 0.27, 0.35, 0.52, 0.70, 1.06, 2.33, and 4.18 mGy, respectively. Median P-KA values were 0.10, 0.26, 0.14, 0.17, 0.77, 0.68, 0.81, and 1.11 Gy cm(2), respectively. The estimated effective dose from ESAK and P-KA values yielded comparable results. The comparison revealed that the ESAK and P-KA values fell far below the reported in the literature. The results showed that the information of the DICOM deader is valuable for dosimetry and optimization.
机译:在这项研究中,我们寻求估计在大型医院中进行数字射线照相X射线检查中的患者辐射剂量。患者暴露因子和KERMA区域产品(P-KA)通过医学(DICOM)标题中的数字成像和547名患者进行回顾性记录。从X射线管输出和记录的曝光因子的测量结果估计入口表面空气Kerma(ESAK),以及从显示P-Ka的控制台作为替代方法。使用适当的转换系数从Esak和P-Ka值估计有效剂量。在胸部PA,胸部LAT,颈椎AP,颈椎脊柱,腹部AP,骨盆AP,腰椎AP和腰椎LAT,中位数ESAK(MGY)被发现为0.13,0.27,0.35,0.52,0.70 ,1.06,2.33和4.18 MGY。中值P-Ka值分别为0.10,0.26,0.14,0.17,0.77,0.68,0.11和1.11 Gycm(2)。来自Esak和P-Ka值的估计有效剂量产生了可比的结果。比较透露,Esak和P-Ka值远远低于文献中报道的。结果表明,DICOM定义的信息对于剂量测定和优化是有价值的。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号