首页> 外文期刊>Journal of clinical laboratory analysis. >Comparative analysis for strength serum sodium and potassium in three different methods: Flame photometry, ion‐selective electrode ( ISE ISE ) and colorimetric enzymatic
【24h】

Comparative analysis for strength serum sodium and potassium in three different methods: Flame photometry, ion‐selective electrode ( ISE ISE ) and colorimetric enzymatic

机译:三种不同方法强度血清钠和钾的比较分析:火焰测光,离子选择性电极(ISE ISE)和比色酶促

获取原文
获取原文并翻译 | 示例
       

摘要

Background Hydroelectrolytic disorders are common in clinical situations and may be harmful to the patient, especially those involving plasma sodium and potassium dosages. Among the possible methods for the dosages are flame photometry, ion‐selective electrode ( ISE ) and colorimetric enzymatic method. Methods We analyzed 175 samples in the three different methods cited from patients attending the laboratory of the University Hospital of the Federal University of Juiz de Fora. The values obtained were statistically treated using SPSS 19.0 software. The present study aims to evaluate the impact of the use of these different methods in the determination of plasma sodium and potassium. Results The averages obtained for sodium and potassium measurements by flame photometry were similar ( P ??.05) to the means obtained for the two electrolytes by ISE . The averages obtained by the colorimetric enzymatic method presented statistical difference in relation to ISE , both for sodium and potassium. In the correlation analysis, both flame photometry and colorimetric enzymatic showed a strong correlation with the ISE method for both dosages. Conclusion At the first time in the same work sodium and potassium were analyzed by three different methods and the results allowed us to conclude that the methods showed a positive and strong correlation, and can be applied in the clinical routine.
机译:背景技术水力电解障碍在临床情况下是常见的,并且可能对患者有害,尤其是涉及血浆钠和钾剂量的患者。在剂量的可能方法中是火焰光度测量,离子选择性电极(ISE)和比色酶法。方法分析了从参加联邦Juiz de Fora大学医院实验室的患者中引用的三种不同方法中的175个样本。使用SPSS 19.0软件进行统计处理所获得的值。本研究旨在评估使用这些不同方法在测定血浆钠和钾的影响。结果对火焰光度测量的钠和钾测量得到的平均值相似(p?&β。05),通过ISE为两种电解质获得的装置。通过比色酶促方法获得的平均值呈现与ISE的统计差异,均用于钠和钾。在相关性分析中,火焰测光和比色酶促与两个剂量的ISE方法表现出强烈的相关性。结论在相同的工作中首次采用三种不同的方法分析钠和钾,结果允许我们得出结论,该方法表现出积极和强的相关性,并且可以应用于临床常规中。

著录项

相似文献

  • 外文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号