摘要

A 72-year-old man with dilated cardiomyopathy and sustained ventricular tachycardia was treated with amiodarone. He visited another hospital because of loss of consciousness. Electrocardiography showed 2: 1 atrioventricular block. Ambulatory electrocardiography showed total heart beats were 59,700 per day. He was referred to our hospital to evaluate his heart. Several types of ventricular tachycardia and ventricular fibrillation were induced by program stimulation during the electrophysiological study. Therefore, an implantable cardioverter-defibrillator was introduced. During defibrillation threshold tests, ventricular fibrillation could not be terminated by the maximal output of 31J. Despite changing the polarity and lead position, stable defibrillation could not be obtained. Finally, successful defibrillation could only be achieved by disconnection of the superior vena cava electrode.
机译:一名72 岁的 男子 与 扩张型心肌病 和 持续性室 性心动过速 与 胺碘酮 治疗。 他访问了 ,因为 意识丧失 的 另一家医院 。 心电图显示 2:1 房室传导阻滞。 动态心电图 显示 总 心脏搏动 是 59700 每天 。 他 提到我们 医院 来评价 他的心脏 。 几种类型的 室性心动过速 和心室颤动 的 电生理研究 过程中 诱导 程序 刺激。 因此 ,引入了 植入式心脏 转复除颤器 。 在 除颤 阈值测试 , 心室纤维性颤动 不能被 31J 的最大 输出 终止。 尽管 改变极性 和 领先地位 , 无法获得 稳定的 除颤。 最后,成功的 除颤 只能由 上腔静脉 电极 的 断开 来实现的。