首页> 外文期刊>Journal of Bionanoscience >Nanostethoscopy: Atomic Force Microscopy Probe Contact Force versus Measured Amplitude of Cardiomyocytic Contractions
【24h】

Nanostethoscopy: Atomic Force Microscopy Probe Contact Force versus Measured Amplitude of Cardiomyocytic Contractions

机译:纳米静脉检查室:原子力显微镜探针接触力与有型心肌收缩的测量幅度

获取原文
获取原文并翻译 | 示例
           

摘要

Cardiomyocytes are one of the most active cells in the human body. Understanding their mechanical contractions via nanostethoscopy has given us better insight into the subtleties of contractions. There is extensive data available on the temporal behavior of contraction in a cardiomyocytes but little is known about how contact force of the AFM probe influences the measurement force of contractions i.e., mean beat force. This investigation suggests that in order to accurately measure mean beat force nanostethoscopy data should be analyzed from the lowest to the highest contact force. In this study optimal contact force was varied from 10 nN to 500 nN and it was observed that the maximum contraction i.e., mean beat force was highest at contact force of 250 nN.
机译:心肌细胞是人体中最活跃的细胞之一。 通过纳米镜检查理解它们的机械收缩使我们能够更好地了解收缩的微妙之处。 在心肌细胞中收缩的时间行为有广泛的数据,但是关于AFM探针的接触力如何影响收缩的测量力,甚至令人满意。,均值搏动力。 该研究表明,为了准确测量平均节拍力纳米镜检查数据,应从最低到最高接触力分析。 在该研究中,最佳接触力从10 nn至500nn变化,并且观察到最大收缩,即250 nn接触力的平均节拍力最高。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号