首页> 外文期刊>Journal of biomechanical engineering. >Methods for Post Hoc Quantitative Computed Tomography Bone Density Calibration: Phantom-Only and Regression
【24h】

Methods for Post Hoc Quantitative Computed Tomography Bone Density Calibration: Phantom-Only and Regression

机译:后HOC定量计算断层扫描骨密度校准方法:仅限幻影和回归

获取原文
获取原文并翻译 | 示例
           

摘要

Quantitative computed tomography (qCT) relies on calibrated bone mineral density data. If a calibration phantom is absent from the CT scan, post hoc calibration becomes necessary. Scanning a calibration phantom after-the-fact and applying that calibration to uncalibrated scans has been used previously. Alternatively, the estimated density is known to vary with CT settings, suggesting that it may be possible to predict the calibration terms using CT settings. This study compares a novel CT setting regression method for post hoc calibration to standard and post hoc phantom-only calibrations. Five cadaveric upper limbs were scanned at 11 combinations of peak tube voltage and current (80-140 kV and 100-300 mA) with two calibration phantoms. Density calibrations were performed for the cadaver scans, and scans of the phantoms alone. Stepwise linear regression determined if the calibration equation terms were predictable using peak tube voltage and current. Peak tube voltage, but not current, was significantly correlated with regression calibration terms. Calibration equation slope was significantly related to the type of phantom (p 0.001), calibration method (p = 0.026), and peak tube voltage (p 0.001), but not current (p = 1.000). The calibration equation vertical intercept was significantly related to the type of phantom (p 0.001 ), and peak tube voltage (p = 0.006), but not calibration method (p = 0.682), or current (p = 0.822). Accordingly, regression can correlate peak tube voltage with density calibration terms. Suggesting that, while standard qCT
机译:定量计算断层扫描(QCT)依赖于校准的骨矿物密度数据。如果CT扫描不存在校准幻像,则HOC校准后可能是必要的。扫描校准幻像之后,先前已经使用了对未校准扫描的校准。或者,已知估计的密度随CT设置而变化,表明可以使用CT设置预测校准术语。本研究比较了新的CT设置回归方法,用于将HOC校准后的标准校准和仅底栖幻像校准。用两个校准模拟的峰值管电压和电流(80-140kV和100-300 mA)的11个组合扫描五个尸体上肢。对尸体扫描进行密度校准,单独扫描幽灵。如果使用峰值管电压和电流可预测校准方程项,则确定逐步线性回归。峰值管电压,但不是电流,与回归校准术语显着相关。校准方程斜率与幻像(P <0.001)的类型显着相关,校准方法(P = 0.026)和峰值管电压(P <0.001),但不是电流(P = 1.000)。校准方程垂直截距与幻影(P <0.001)的类型显着相关,峰值管电压(P = 0.006),但不校准方法(P = 0.682)或电流(P = 0.822)。因此,回归可以将峰值管电压与密度校准术语相关联。建议,虽然标准QCT

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号