Use of perfluorocarbon compound in the endorectal coil to improve MR spectroscopy of the prostate.

机译:在直肠内线圈中使用全氟化碳化合物以改善前列腺的MR光谱。

获取原文
获取原文并翻译 | 示例
       

摘要

OBJECTIVE: The purpose of this study was to evaluate the utility of perfluorocarbon (PFC) compound compared with air in the endorectal coil in improving the quality of conventional MR spectroscopy of the prostate. SUBJECTS AND METHODS: A total of 62 consecutively registered patients were selected. MR spectroscopy of the prostate was performed with PFC in the endorectal coil for 34 of the patients and with air for 28. In the cases of 13 of the 28 patients, MR spectroscopy was repeated with a PFC-filled endorectal coil. The spectral line widths and the spectral separations from MR spectroscopy with an air-filled endorectal coil were compared with those obtained with a PFC-filled coil. RESULTS: In all 62 patients, the mean line width values were reduced, from 13.3 +/- 3.0 Hz with air to 7.3 +/- 2.0 Hz with PFC (p = 0.0001, Student's t test). In 13 patients who underwent MR spectroscopy with air followed by MR spectroscopy with PFC, the mean line width values were reduced, from 14.8 +/- 3.4 Hz with air to 7.0 +/- 1.5 Hz with PFC (p = 0.0001, Student's t test). In 72 voxels analyzed for spectral separation, clear separations between the choline, polyamine, and creatine peaks were found in 57 voxels with PFC and four with air. Better splitting of the citrate peaks was found in 35 voxels with PFC and one with air. CONCLUSION: The use of PFC compound in the endorectal coil is a safe and cost-effective way to consistently generate high-quality prostate MR spectroscopic results and to substantially improve quantitation of prostatic metabolites. These improvements should increase the diagnostic value of MR spectroscopy in the care of patients with prostate cancer.
机译:目的:本研究的目的是评估全氟化碳(PFC)化合物与直肠内线圈中的空气相比,在提高常规前列腺MR光谱学质量方面的实用性。受试者和方法:总共选择了62名连续登记的患者。前列腺磁共振波谱是在34例患者的直肠内线圈中使用PFC进行的,其中28例是在空气中进行的。在28例患者中的13例中,使用充满PFC的直肠内线圈重复进行MR光谱。将空气填充的直肠内线圈与MR光谱的谱线宽度和光谱间隔进行了比较,结果与使用PFC填充的线圈获得的谱线宽度和光谱间隔进行了比较。结果:在所有62例患者中,平均线宽值均从空气中的13.3 +/- 3.0 Hz降低到PFC时的7.3 +/- 2.0 Hz(p = 0.0001,Student's t检验)。在13例接受空气MR光谱检查后再进行PFC MR光谱检查的患者中,平均线宽值从使用空气的14.8 +/- 3.4 Hz降低到使用PFC的7.0 +/- 1.5 Hz(p = 0.0001,Student's t检验)。在分析光谱分离的72个体素中,在57个使用PFC的体素和四个使用空气的体素中发现了胆碱,多胺和肌酸峰之间的清晰分离。在35个使用PFC的体素和一个使用空气的体素中,发现柠檬酸盐峰的拆分更好。结论:在直肠内线圈中使用PFC化合物是一种安全且具有成本效益的方法,可以始终如一地产生高质量的前列腺MR光谱结果,并显着改善前列腺代谢产物的定量。这些改进应提高MR光谱在前列腺癌患者护理中的诊断价值。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号