Toward a Standardization of Ultrasound Scanners for Dynamic Contrast-Enhanced Ultrasonography: Methodology and Phantoms

机译:朝着动态对比度增强超声检查的超声扫描仪的标准化:方法和幻影

获取原文
获取原文并翻译 | 示例
           

摘要

Abstract The standardization of ultrasound scanners for dynamic contrast-enhanced ultrasonography (DCE-US) is mandatory for evaluation of clinical multicenter studies. We propose a robust method using a phantom for measuring the variation of the harmonic signal intensity obtained from the area under the time-intensity curve versus various contrast-agent concentrations. The slope of this measured curve is the calibration parameter. We tested our method on two devices from the same manufacturer (AplioXV and Aplio500, Toshiba, Tokyo, Japan) using the same settings as defined for a French multicenter study. The Aplio500's settings were adjusted to match the slopes of the AplioXV, resulting in the following settings on the Aplio500: at 3.5?MHz: MI?=?0.15; CG?=?35?dB and at 8?MHz: MI?=?0.10; CG?=?32?dB. This calibration method is very important for future DCE-US multicenter studies. ]]>
机译:摘要对于动态对比增强超声检查(DCE-US)的超声扫描仪的标准化是用于评估临床多中心研究的强制性。 我们提出了一种使用幻像来测量从时间强度曲线的区域获得的谐波信号强度的变化的鲁棒方法,而是各种对比剂浓度。 该测量曲线的斜率是校准参数。 我们在同一制造商(APLIOXV和APLIO500,TOSHICA,TOKYO,日本)的两个设备上测试了我们的方法,使用与法国多中心研究定义的相同的设置。 调整APLIO500的设置以匹配APLIOXV的斜率,从而导致APLIO500上的以下设置:3.5?MHz:MI?=?0.15; cg?=?35?db和8?mhz:mi?=?0.10; cg?=?32?db。 这种校准方法对于未来的DCE-US多中心研究非常重要。 ]]>