Optimizing radiation dose parameters in MDCT arthrography of the shoulder: illustration of basic concepts in a cadaveric study

机译:优化肩部MDCT散射术中的辐射剂量参数:尸体研究中基本概念的插图

获取原文
获取原文并翻译 | 示例
           

摘要

ObjectiveTo determine in a cadaveric study the lowest achievable radiation dose and optimal tube potential generating diagnostic image quality in multidetector computed tomography (MDCT) arthrography of the shoulder.Materials and methodsSix shoulders from three human cadavers were scanned using a 256-MDCT system after intra-articular injection of diluted iodinated contrast material. Using six decreasing radiation dose levels (CTDIvol: 20, 15, 10, 8, 6, and 4mGy) and for each dose level, four decreasing tube potentials (140, 120, 100, and 80kVp), image noise and contrast-to-noise ratio (CNR) were measured. Two independent and blinded observers assessed the overall diagnostic image quality, subjective amount of noise, and severity of artifacts according to a four-point scale. Influence of those MDCT data acquisition parameters on objective and subjective image quality was analyzed using the Kruskal-Wallis and Wilcoxon signed-rank tests, and pairwise comparisons were performed.ResultsMultidetector CT protocols with radiation doses of 15mGy or higher, combined with tube potentials of 100kVp or higher, were equivalent in CNR to the reference 20mGy-140kVp protocol (all p0.054). Above a CTDIvol of 10mGy and a tube potential of 120kVp, all protocols generated diagnostic image quality and subjective noise equivalent to the 20mGy-140kVp protocol (all p0.22).ConclusionsDiagnostic image quality in MDCT arthrography of the shoulder can be obtained with a radiation dose of 10mGy at an optimal tube potential of 120kVp, corresponding to a reduction of up to 50% compared with standard-dose protocols, and as high as 500% compared with reported protocols in the literature.
机译:ObjectiveTo在尸体研究中确定最低可实现的辐射剂量和最佳管电位在肩部的多票计算机断层扫描(MDCT)射突中产生诊断图像质量。在内部 - 关节注射稀释的碘化造影剂。使用六次减少辐射剂量水平(CTDIVOL:20,15,10,8,6,6和4mgy),以及每个剂量水平,四个减小管电位(140,120,100和80kVP),图像噪声和对比度测量噪声比(CNR)。两个独立和盲目的观察者根据四点比例评估了整体诊断图像质量,主观噪声量和伪影的严重程度。使用Kruskal-Wallis和Wilcoxon签名 - 秩检验分析了这些MDCT数据采集参数对客观和主观图像质量的影响,并进行成对比较。辐射剂量为15强或更高的辐射剂量,与100kvp的管电位相结合的细胞比较。或更高,在CNR中等同于参考20MGY-140KVP协议(所有P0.054)。高于CTDIVOL的10MGY和120kVP的管电位,所有协议产生诊断图像质量和相当于20MGY-140KVP协议(所有P0.22)的主观噪声在120kVP的最佳管电位下10mgy的剂量,与标准剂量方案相比减少至多50%,与文献中报告的议定书相比高达500%。

著录项

  • 来源
    《Skeletal radiology》 |2019年第8期|共8页
  • 作者单位

    Lausanne Univ Hosp Dept Diagnost &

    Intervent Radiol Rue Bugnon 46 CH-1011 Lausanne Switzerland;

    Lausanne Univ Hosp Dept Diagnost &

    Intervent Radiol Rue Bugnon 46 CH-1011 Lausanne Switzerland;

    Lausanne Univ Hosp Dept Diagnost &

    Intervent Radiol Rue Bugnon 46 CH-1011 Lausanne Switzerland;

    Philips Healthcare Rue Deux Gares 80 B-1070 Brussels Belgium;

    Catholic Univ Louvain Clin Univ St Luc Dept Radiol &

    Med Imaging Ave Hippocrate 10 B-1200;

    Lausanne Univ Hosp Dept Diagnost &

    Intervent Radiol Rue Bugnon 46 CH-1011 Lausanne Switzerland;

  • 收录信息
  • 原文格式 PDF
  • 正文语种 eng
  • 中图分类 放射医学;
  • 关键词

    MDCT; CT; Shoulder; Arthrography; Radiation dose; Image quality;

    机译:MDCT;CT;肩部;棱锥形;辐射剂量;图像质量;

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号