Analytical Assessment of the New Roche Cobas t 711 Fully Automated Coagulation Analyzer

机译:新型罗氏COBAS T 711全自动凝固分析仪的分析评估

获取原文
获取原文并翻译 | 示例
           

摘要

This study aimed to provide a preliminary evaluation of the analytical performance of the new Roche cobas t 711 fully automated coagulation analyzer, which uses both liquid and lyophilized reagent cassettes. The analytical assessment included analysis of imprecision and linearity of prothrombin time (PT), activated partial thromboplastin time (APTT), and fibrinogen on cobas t 711 analyzer. Test results of 120 routine plasma samples were also compared with those obtained using two other coagulation analyzers (Instrumentation Laboratory ACL TOP 700 and Stago STA-R MAX). The accuracy, imprecision, and comparability of manual and automatic lyophilized material resuspension were also evaluated using 200 routine plasma samples. Overall, automatic resuspension was found to be more precise than, and equally accurate as, manual reconstitution, with coefficient of variations (CV%) three- to sixfold lower compared with manual reconstitution. The analytical imprecision was found to be excellent, as attested by total CV% of 0.7% for PT, 1.7 to 1.8% for APTT, and 1.9 to 3.2% for fibrinogen. Linearity was excellent over a clinically significant range of PT, APTT, and fibrinogen values, displaying correlation coefficients comprised between 0.994 and 0.999. Methods comparison studies revealed that results of PT, APTT, and fibrinogen on cobas t 711 are globally aligned with those obtained using identical plasma samples on IL ACL TOP 700 and Stago STA-R MAX, displaying correlation coefficients of 0.97 for PT, 0.81 and 0.88 for APTT, 0.90 and 0.94 for fibrinogen, respectively. In conclusion, the results of this preliminary evaluation demonstrate that PT, APTT, and fibrinogen on cobas t 711 coagulation analyzer displays excellent performance for routine use in clinical laboratories.
机译:本研究旨在提供新的Roche Cobas T 711全自动凝固分析仪的分析性能的初步评估,它使用液体和冻干试剂盒。分析评估包括凝血酶原时间(PT)的不精确和线性度的分析,活化的部分凝血形成素时间(APTT)和COBAS T 711分析仪的纤维蛋白原。还与使用另外两种凝固分析仪(仪表实验室ACL Top 700和STAGO STA-R Max)的测试结果相比,还将120例常规等离子体样品进行比较。还使用200例常规等离子体样品评估手动和自动冻干材料重悬浮的准确性,不精确和可比性。总的来说,发现自动重悬浮于比手动重构更精确,并且同样准确,与手动重构相比,具有变化系数(CV%)三到六倍。发现分析不精确是优异的,对于Pt的总CV%的CV%,对于脂蛋白原的Pt,1.7%至1.8%,1.9%至3.2%。线性度在临床上显着的Pt,aptt和纤维蛋白原值范围内具有优异的,显示相关系数在0.994和0.999之间。方法对比较研究表明,COBAS T 711上的PT,APTT和纤维蛋白原的结果与在IL ACL顶部700和STAGO STA-R MAX上使用相同的等离子体样品获得的那些,显示为PT,0.81和0.88的相关系数为0.97对于纤维蛋白原的APTT,0.90和0.94分别。总之,该初步评价结果表明,COBAS T 711凝血分析仪上的PT,APTT和纤维蛋白原对临床实验室的常规使用表现出优异的性能。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号