Optimisation of direct digital chest radiography using Cu filtration

机译:CU过滤的直接数字胸部射线照相的优化

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

Purpose: To assess the impact of additional Cu filtration with the DRX-1 detector in optimisation of adult chest radiography. Methods: A simulated anatomical phantom was imaged at different levels of Cu filtration (0 mm, 0.1 mm, 0.3 mm and 0.5 mm of Cu) at 120 kVp. Image quality was assessed using exposure index, conspicuity index, signal-to-noise ratio and pixel density. Dose area product (DAP) meter was used to measure dose; effective dose and risk of exposure induced cancer death (REID) were estimated using the PCXMC software 2.0. Results: Exposure index increased with increasing Cu thickness; 0.3 mm Cu filtration increased conspicuity index of simulated nodules, with a minimal decrease in signal-to-noise ratio, and pronounced decrease in pixel density. Patient doses were reduced by 20.7%, 36.6%, and 39.79% at 0.1 mm, 0.3 mm, and 0.5 mm levels of Cu filtration respectively. Conclusion: Additional beam filtration using 0.3 mm Cu filter with the DXR-1 detector reduced patient doses by 37% without decline in image quality.
机译:目的:评估额外Cu过滤与DRX-1探测器的影响,以优化成人胸部射线照相。方法:在120 kVp下在不同水平的Cu过滤水平(0mm,0.1mm,0.3mm和0.5mm)的不同水平上成像模拟的解剖学体验。使用曝光指数,链接性指数,信噪比和像素密度来评估图像质量。剂量区域产品(DAP)仪表用于测量剂量;使用PCXMC软件2.0估计有效剂量和暴露诱导的癌症死亡(Reid)的风险。结果:曝光指数随着Cu厚度的增加而增加; 0.3mm Cu过滤增加了模拟结节的光纤指数,发信噪比的最小降低,并发出了像素密度的显着降低。患者剂量分别降低20.7%,36.6%和39.79%,分别为0.1mm,0.3mm和0.5mm水平的Cu过滤。结论:使用0.3mM Cu过滤器的额外光束过滤,DXR-1检测器减少37%,患者剂量减少37%,没有图像质量下降。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号