The spatial non-uniformity of a newly developed electrically calibrated radiometer system is characterized. The detector is a thin-film thermopile with an integral electrical heater. This paper shows that the local responsivities of these detectors have a rotational symmetry and a quadratic dependence on the distance to the centre of the detector. Typical measurement errors, which can be caused by applying different beam sizes, different power distributions of the beams, and different beam positions impinging on the detector, are investigated. A significant reduction of measurement uncertainties is achieved using specific corrections. The influence of spatial non-uniformity on responsivity measurements is experimentally checked with various types of beam.
机译:特征在于新开发的电校准辐射计系统的空间不均匀性。 探测器是具有整体电加热器的薄膜热电堆。 本文表明,这些探测器的局部响应具有旋转对称性和对探测器中心的距离的二次依赖性。 研究了典型的测量误差,其可以通过施加不同的光束尺寸,光束的不同功率分布以及撞击检测器上的不同光束位置而引起的。 使用特定校正实现测量不确定性的显着降低。 通过各种类型的光束实验检查空间不均匀性对响应度测量的影响。