首页> 外文期刊>Materiale Plastice >Comparison of Classic and Metal Reinforced Maxillary Acrylic Complete Dentures Fracture Resistance
【24h】

Comparison of Classic and Metal Reinforced Maxillary Acrylic Complete Dentures Fracture Resistance

机译:经典和金属增强上颌丙烯酸完全假牙裂缝抗性的比较

获取原文
获取原文并翻译 | 示例
           

摘要

The aim of our study was to evaluate fracture resistance of the PMMA complete dentures and to make a comparison between the classic acrylic maxillary complete dentures and the metal-reinforced ones, in terms of fracture resistance during function and to compare the fracture pattern of the two groups of complete dentures. We used maxillary complete dentures made of heat-processed acrylate, 5 exclusively and 5 in which a metal-cast net in the form of a mesh was inserted after being casted, worn by patients for a period of 5 years. A bilateral progressive force was applied to the PM2-M1 area until the samples failed, recording the force value at which the first change occurred. A Universal Loading Machine was used. The results were analyzed using the ANOVA method. For the classic acrylic complete dentures group, the minimum fracture force value was 1060 N, the maximum fracture force value being 2443 N. For the reinforced complete dentures group, the minimum fracture force value was 3320 N, the maximum value being 3760 N. Statistical analysis of data using the ANOVA shows that the results are statistically relevant, with a p value of 0.000137 (p<0,05). All the dentures from the classic acrylic complete dentures group fractured presenting two or more fracture lines, most of them involving also fractures on the acrylic teeth and breaking into multiple pieces, while all the dentures from the reinforced complete dentures group fractured presenting one or two fracture lines, most of them without detachable fragments and involving also fractures on the acrylic teeth.
机译:我们的研究目的是评估PMMA完整假牙的骨折性,并在功能期间的断裂抗性方面进行经典的丙烯酸上颌全部义齿和金属增强物的比较,并比较两者的骨折图案一组完整的假牙。我们使用了由热处加工丙烯酸酯制成的上颌完全义齿,排他性5,其中患者在铸造后,患者穿过5年后,佩戴物形式的金属铸网。将双侧渐进力施加到PM2-M1面积,直到样品失效,记录第一次变化发生的力值。使用通用装载机。使用ANOVA方法分析结果。对于经典的丙烯酸类完整假牙组,最小断裂力值为1060 n,最大断裂力值为2443 n。对于加强的完整假牙组,最小骨折力值为3320 n,最大值为3760 n。统计使用ANOVA的数据分析表明,结果在统计上相关,AP值为0.000137(P <0,05)。来自经典的丙烯酸完整假牙组的所有假牙骨折呈现两条或更多个骨折线,其中大部分涉及丙烯酸牙齿上的骨折并分成多个碎片,而来自增强的完整假牙组的所有假牙裂缝呈现一个或两个骨折线,其中大部分都没有可拆卸碎片,也涉及丙烯酸牙齿上的骨折。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号