Lenstar (R) LS 900 vs Pentacam (R)-AXL: Comparative study of ocular biometric measurements and intraocular lens power calculation

机译:LENSTAR(R)LS 900 VS PENTACAM(R)-AXL:眼镜生物测量和眼内透镜功率计算的比较研究

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

Purpose: Comparison of biometric measurements and calculation of intraocular lens with a new biometer (Pentacam (R)-AXL, Oculus, Germany) and a reference biometer (Lenstar LS 900 (R), Haag-Streit AG, Switzerland), in order to assess the agreement between these two devices. Setting: Centro Hospitalar de Leiria, Portugal Materials and methods: Prospective, institutional study, in which measurements of axial length, anterior chamber depth from the corneal epithelium and endothelium to the anterior surface of the lens (anterior chamber depth ext and anterior chamber depth int), central corneal thickness and keratometry readings of the flattest and steepest meridians (K1 and K2) were obtained with the two systems. Intraocular lens calculation was also performed, using the Haigis, SRK/T and HofferQ formulas. Results: The study sample included 136 eyes of 79 patients. Of these, 42 were women and 37 were men. Statistically significant differences were found (p 0.05, paired T test) in K1, K2 and central corneal thickness between the 2 biometers. Intraocular lens calculation with the Lenstar (R) and the Pentacam (R) with Haigis, SRK/T and HofferQ formulas showed statistically significant differences (p 0.05 Paired T test). Conclusion: Axial length measurements obtained with the Pentacam (R) and Lenstar (R) appear to be interchangeable, while measurements of anterior chamber depth, K1 and K2, and central corneal thickness do not appear to be interchangeable between different devices. Statistically significant differences were found in the calculation of intraocular lenses in all formulas used.
机译:目的:具有新的生物镜的生物识别测量和计算的生物晶状体(Pentacam(R)-AXL,Oculus,Oculus,德国)和参考生物镜(Lenstar LS 900(R),Haag-Streit AG,瑞士)进行比较,以便评估这两个设备之间的协议。环境:Centro Haparalar de Leiria,葡萄牙材料和方法:前瞻性,制度研究,其中轴向长度测量,从角膜上皮和内皮的前表面深度到透镜前表面(前房深度探测器和前房深度int。 ),用两种系统获得最典型和最陡的经络(K1和K2)的中央角膜厚度和静脉测定读数。使用HAIGIS,SRK / T和Hofferq公式还进行眼内晶状体计算。结果:研究样品包括136只7例患者。其中,42个是女性,37个是男性。发现统计学上的显着差异(P& 0.05,配对T测试)在2个生物夹之间的K1,K2和中央角膜厚度中。用LENSTAR(R)和HAIGIS,SRK / T和Hofferq公式的子宫内晶状体计算显示出统计学上显着的差异(P <0.05对T测试)。结论:用戊肟(R)和Lenstar(R)获得的轴向长度测量似乎是可互换的,同时在不同装置之间的可互换腔室深度,K1和K2和中央角膜厚度的测量。在所有公式中的眼内晶状体计算中发现了统计上显着的差异。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号