首页> 外文期刊>Eye & contact lens >Repeatability and Agreement of a Swept-Source Optical Coherence Tomography-Based Biometer IOLMaster 700 Versus a Scheimpflug Imaging-Based Biometer AL-Scan in Cataract Patients
【24h】

Repeatability and Agreement of a Swept-Source Optical Coherence Tomography-Based Biometer IOLMaster 700 Versus a Scheimpflug Imaging-Based Biometer AL-Scan in Cataract Patients

机译:基于扫描源光学相干性断层扫描的生物计的重复性和协议与白内障患者的Scheimpflum成像的生物计al-scan

获取原文
获取原文并翻译 | 示例
           

摘要

Purpose: To compare the repeatability and agreement between a swept-source biometer and a Scheimpflug biometer in cataract patients. Methods: Three consecutive measurements were obtained using a swept-source biometer (IOLMaster 700) and a Scheimpflug biometer (AL-Scan) in 52 eyes of 52 patients. Keratometry, central corneal thickness (CCT), anterior chamber depth (ACD), axial length, and white-to-white (WTW) distance were recorded. Astigmatism values were transformed into vector components of J0 and J45. Intraoperator repeatability was analyzed using intraclass correlation coefficients (ICCs) and reproducibility coefficients (RCs). Agreement of measurements between the two devices was evaluated using the Bland-Altman method. Results: The IOLMaster 700 showed higher ICCs and lower RCs for the mean keratometry (Km) (P <= 0.018), CCT (P <= 0.027), and ACD (P <= 0.001) measurements, whereas the AL-Scan showed higher ICC and lower RC for the J45 vector component of astigmatism at the 2.4-mm zone (P <= 0.034). Both the devices had excellent repeatability (ICC=0.999) in axial length measurement. Systematic differences were found in Km, CCT, ACD, and WTW (P <= 0.018) between the devices. The mean difference for Km was -0.196 and -0.144 D measured at the 2.4-mm zone and 3.3-mm zone, respectively. The corresponding mean difference for CCT, ACD, and WTW distance was 14.92 mu m, -0.017 mm, and 0.283 mm, respectively. These differences led to a statistically significant but clinically insignificant difference in the prediction of intraocular lens power. Conclusions: This study showed significant differences in anterior segment measurement repeatability and agreement between a swept-source biometer and a Scheimpflug biometer in eyes with cataract.
机译:目的:比较白内障患者中扫掠源生物差管和Scheimpflug生物计之间的重复性和协议。方法:使用52名患者的52只眼中的Scheimpflug生物计(IOLMASTER 700)和Scheimpflug生物仪(Al-Scal)获得三次连续测量。记录,中央角膜厚度(CCT),前房深度(ACD),轴向长度和白色(WTW)距离。散光值转化为J0和J45的载体组分。使用脑内相关系数(ICC)和再现性系数(RCS)分析史内控器可重复性。使用Bland-Altman方法评估两种设备之间的测量值。结果:IOLMASTER 700显示出更高的ICC和较低的RC,用于平均静脉测量(KM)(P <= 0.018),CCT(P <= 0.027)和ACD(P <= 0.001)测量,而AL扫描显示得更高ICC和下RC为2.4mm区的散光的J45矢量分量(P <= 0.034)。这些设备在轴向长度测量中具有出色的重复性(ICC = 0.999)。在设备之间,在KM,CCT,ACD和WTW(P <= 0.018)中发现了系统差异。 KM的平均差异分别在2.4毫米区域和3.3mm区测量的-0.196和-0.144d。 CCT,ACD和WTW距离的相应平均差异分别为14.92μm,-0.017mm和0.283mm。这些差异导致了统计上显着但临床上微不足道的差异,在眼内透镜功率预测。结论:本研究表明,扫掠源生物仪和白内障的眼睛中的Scheimpflug病差之间存在显着差异。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号