Objectives Detectability experiments performed to assess the diagnostic performance of computed tomography (CT) images should represent the clinical situation realistically. The purpose was to develop anatomically realistic phantoms with low-contrast lesions for detectability experiments. Methods Low-contrast lesions were digitally inserted into a neck CT image of a patient. The original and the manipulated CT images were used to create five phantoms: four phantoms with lesions of 10, 20, 30, and 40 HU contrast and one phantom without any lesion. Radiopaque 3D printing with potassium-iodide-doped ink (600 mg/mL) was used. The phantoms were scanned with different CT settings. Lesion contrast was analyzed using HU measurement. A 2-alternative forced choice experiment was performed with seven radiologists to study the impact of lesion contrast on detection accuracy and reader confidence (1 = lowest, 5 = highest). Results The phantoms reproduced patient size, shape, and anatomy. Mean +/- SD contrast values of the low-contrast lesions were 9.7 +/- 1.2, 18.2 +/- 2, 30.2 +/- 2.7, and 37.7 +/- 3.1 HU for the 10, 20, 30, and 40 HU contrast lesions, respectively. Mean +/- SD detection accuracy and confidence values were not significantly different for 10 and 20 HU lesion contrast (82.1 +/- 6.3% vs. 83.9 +/- 9.4%, p = 0.863 and 1.7 +/- 0.4 vs. 1.8 +/- 0.5, p = 0.159). They increased to 95 +/- 5.7% and 2.6 +/- 0.7 for 30 HU lesion contrast and 99.5 +/- 0.9% and 3.8 +/- 0.7 for 40 HU lesion contrast (p < 0.005). Conclusions A CT image was manipulated to produce anatomically realistic phantoms for low-contrast detectability experiments. The phantoms and our initial experiments provide a groundwork for the assessment of CT image quality in a clinical context.
机译:目标检测性实验评估计算断层扫描(CT)图像的诊断性能应逼真地代表临床局势。目的是通过用于可检测性实验的低对比度病变进行解剖学上现实的幽灵。方法将低对比度病变数字纳入患者的颈部CT图像中。原始和操纵的CT图像用于产生五种幻影:四分之一,病变为10,20,30和40 HU对比度和一个幻象而没有任何病变。使用含钾碘化物(600mg / ml)的无线电桥3D印刷。用不同的CT设置扫描幽灵。使用HU测量分析病变对比度。使用七位放射科学家进行了一个2替代的强制选择实验,以研究病变对比对检测精度和读者置信度的影响(1 =最低,5 =最高)。结果幻影再现患者大小,形状和解剖学。平均低对比度病变的+/-SD对比度值为9.7 +/- 1.2,18.2 +/- 2,30.2 +/- 2.7,以及10,20,30和40 Hu的37.7 +/- 3.1胡锦涛对比病变分别。平均+/- SD检测精度和置信度对10和20的胡病灶造影没有显着差异(82.1 +/- 6.3%,+ 0.863和1.7 +/- 0.4与1.8 + / - 0.5,p = 0.159)。 30 Hu病灶对比度和2.6 +/- 0.7增加到95 +/- 0.7的95 +/- 0.7%和3.8 +/- 0.7,适用于40例病灶对比度(P <0.005)。结论被操纵CT图像以产生解剖学上的幽灵,用于低对比度可检测性实验。幽灵和我们的初始实验为在临床背景下评估CT图像质量提供了基础。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号