ABSTRACT Background: Elastomeric infusion pumps are widely used in the delivery of parenteral medications in the home, but real-life conditions may not match calibration or standardised testing conditions. This study investigated the impact of changes in infusion pump height and/or back pressure on infusion pump function. Methods: Volume delivered after one day, infusion duration, average and peak flow rates and time spent within stated accuracy were determined for four elastomeric and one electronic pump using gravimetric technique. Experiments were repeated after altering the height of the pump relative to the output (±40cm, ±20cm) and/or adding a back pressure (10-30mmHg) to the output of an attached catheter. Results: Under ideal operating conditions, the flow rate deviated from that specified by the manufacturer and between 88.5% and 99% of the total infusion volume was delivered. Varying the height or applying back pressure led to further changes in average flow rates and the volume of infusion solution delivered by the elastomeric pumps, but had little effect on the electronic pump. Conclusions: Clinicians should consider potential impact on drug delivery, safety and therapeutic effect for home infusion patients given variations in infusion pump performance observed in this study.
机译:摘要背景:弹性输液泵广泛用于家庭中的肠外药物,但现实生活条件可能与校准或标准化的测试条件不匹配。本研究研究了输液泵高度和/或背压变化对输液泵功能的影响。方法:使用重量技术为四个弹性体和一个电子泵确定一天后,输液持续时间,平均和峰值流速和时间在规定的精度下递送的体积。在改变泵的高度相对于输出(±40cm,±20cm)和/或将背压(10-30mmHg)添加到连接导管的输出之后重复实验。结果:在理想的操作条件下,延误了由制造商规定的流量和88.5%和99%的总输注体积的流量。改变高度或施加背压导致平均流速的进一步变化和弹性泵输送的输注溶液的体积,但对电子泵几乎没有影响。结论:临床医生应考虑对药物输送的潜在影响,家庭输注患者的安全性和治疗效果给出该研究中观察到的输液泵性能的变化。

著录项

相似文献

  • 外文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号