Hyperthermia treatment planning guided applicator selection for sub-superficial head and neck tumors heating

机译:热疗治疗计划引导涂药器选择亚表面头部和颈部肿瘤加热

获取原文
获取原文并翻译 | 示例
           

摘要

Purpose: In this study, we investigated the differences in hyperthermia treatment (HT) quality between treatments applied with different hyperthermia systems for sub-superficial tumours in the head and neck (H&N) region.Materials and methods: In 24 patients, with a clinical target volume (CTV) extending up to 6cm from the surface, we retrospectively analysed the predicted HT quality achievable by two planar applicator arrays or one phased-array hyperthermia system. Hereto, we calculated and compared the specific absorption rate (SAR) and temperature distribution coverage of the CTV and gross tumour volume (GTV) for the Lucite cone applicator (LCA: planar), current sheet applicator (CSA: planar) and the HYPERcollar (phased-array).Results: The HYPERcollar provides better SAR coverage than planar applicators if the target region is fully enclosed by its applicator frame. For targets extending outside the HYPERcollar frame, sufficient SAR coverage (25% target coverage, i.e. TC2575%) can still be achieved using the LCA when the target is fully under the LCA aperture and not deeper than 50mm from the patient surface.Conclusion: Simulations predict that the HYPERcollar (hence also its successor the HYPERcollar3D) is to be preferred over planar applicators such as LCA and current sheet applicator in sub-superficial tumours in the H&N region when used within specifications.
机译:目的:在这项研究中,我们研究了在头部和颈部(H&N)区域的亚表面肿瘤中应用不同热疗系统适用的治疗之间的热疗治疗(HT)质量的差异。材料和方法:24例患者,临床从表面延伸到6厘米的目标体积(CTV),我们回顾性地分析了两个平面涂抹器阵列或一个相位阵列热疗系统可实现的预测HT质量。我们计算并比较了Lucite锥形涂布器(LCA:Planar),当前纸张涂布器(CSA:Placear)和高压杆(相控阵)。如果目标区域完全由其涂抹器框架完全封闭,则高压机提供比平面涂抹器更好的SAR覆盖。对于延伸的目标在高压机框架外,使用LCA仍然可以通过LCA在LCA孔径下方且距离患者表面的50mm更深而不是50mm时仍然可以实现足够的SAR覆盖率(25%的目标覆盖率,即TC2575%)。结论:模拟预测超级型(因此也是其继承者的HyperCollAR3D)是在规格内使用的,在H&N区域中的亚表面肿瘤中的平面施加器(如LCA和当前纸张涂敷器)上。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号