首页> 外文期刊>International journal of analytical chemistry >Direct Analysis of Low-Volatile Molecular Marker Extract from Airborne Particulate Matter Using Sensitivity Correction Method
【24h】

Direct Analysis of Low-Volatile Molecular Marker Extract from Airborne Particulate Matter Using Sensitivity Correction Method

机译:使用灵敏度校正法从空气颗粒物质中直接分析低挥发性分子标记提取物

获取原文
获取原文并翻译 | 示例
           

摘要

Molecular marker analysis of environmental samples often requires time consuming preseparation steps. Here, analysis of low-volatile nonpolar molecular markers (5-6 ring polycyclic aromatic hydrocarbons or PAHs, hopanoids, and n-alkanes) without the preseparation procedure is presented. Analysis of artificial sample extracts was directly conducted by gas chromatography-mass spectrometry (GC-MS). After every sample injection, a standard mixture was also analyzed to make a correction on the variation of instrumental sensitivity caused by the unfavorable matrix contained in the extract. The method was further validated for the PAHs using the NIST standard reference materials (SRMs) and then applied to airborne particulate matter samples. Tests with the SRMs showed that overall our methodology was validated with the uncertainty of ~30%. The measurement results of airborne particulate matter (PM) filter samples showed a strong correlation between the PAHs, implying the contributions from the same emission source. Analysis of size-segregated PM filter samples showed that their size distributions were found to be in the PM smaller than 0.4?μm aerodynamic diameter. The observations were consistent with our expectation of their possible sources. Thus, the method was found to be useful for molecular marker studies.
机译:环境样品的分子标记分析通常需要耗时消耗的预备步骤。这里,提出了不具有预分离程序的低挥发性非极性分子标记(5-6个环多环芳烃或PAH)的分析,没有预留过程。通过气相色谱 - 质谱(GC-MS)直接进行人造样品提取物的分析。在每次样品注射后,还分析了标准混合物,以纠正由提取物中含有的不利基质引起的仪器敏感性的变化。使用NIST标准参考材料(SRMS)进一步验证该方法,然后应用于空气传播的颗粒物质样品。随着SRMS的测试表明,我们的方法总体方法被验证了〜30%的不确定性。空气颗粒物质(PM)过滤样品的测量结果显示PAHS之间的强烈相关,暗示了来自相同排放源的贡献。分析尺寸分离的PM过滤器样品显示,发现它们的尺寸分布在小于0.4ΩΩΩΩ·μm的PM中。观察结果与我们对可能的来源的期望一致。因此,发现该方法可用于分子标记研究。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号