Evaluation of procalcitonin assay on the Abbott Architect i2000 SR (R) analyzer

机译:Abbott Architect I2000 SR(R)分析仪对ProCalcitonin测定的评价

获取原文
获取原文并翻译 | 示例
       

摘要

Procalcitonin (PCT) is a useful biomarker for the diagnosis of bacterial infection. Its measurement is used routinely as a valuable tool for antibiotic treatment decision. The aim of this study is to assess the analytical performance of the new Architect Brahms PCT (R) reagents on the Abbott Architect i2000-SR (R) immuno-analyzer. The Architect PCT (R) assay was evaluated according to the modified SFBC protocol, in accordance with the NF EN ISO 15189 standard. Sixty two samples from patients hospitalized in Necker Hospital (Paris) have been tested with the evaluated method, and results were compared to those of the PCT Kryptor Brahms (R) method. Analytical performances tested complied with those announced by the manufacturer. Linear regression showed a strong correlation between the two methods (r 0.99), despite a tendency for overestimation by the new method (y=1.10 x + 0.05). Bland-Altman plots confirmed this strong correlation by showing only two points outside the acceptable limits without clinical incidence, given the high PCT concentrations (over 10 ng/mL) in those samples. In conclusion, the new PCT assay on the Abbott Architect i2000-SR (R) shows excellent analytical performances, even at low concentrations. A slight positive bias compared to the Brahms Kryptor (R) was observed, but did not lead to inappropriate clinical decisions.
机译:ProCalcitonin(PCT)是一种有用的生物标志物,用于诊断细菌感染。其测量通常是抗生素治疗决策的有价值的工具。本研究的目的是评估Abbott Architect I2000-SR(R)免疫分析仪上的新建筑师Brahms PCT(R)试剂的分析性能。根据“修改的SFBC协议”,根据NF EN ISO 15189标准评估建筑师PCT(R)测定。通过评估的方法测试六十次来自颈部医院(巴黎)的患者的患者,并将结果与​​PCT Kryptor Brahms(R)方法进行了比较。测试的分析表演符合制造商宣布的表演。线性回归在两种方法(R> 0.99)之间存在强烈的相关性,尽管通过新方法估计倾向(Y = 1.10 x + 0.05)。在没有临床发病率的情况下仅显示可接受的限制之外的两个点,鉴于这些样品中的高PCT浓度(超过10ng / ml),Bland-Altman绘图证实了这种强烈的相关性。总之,Abbott Architect I2000-SR(R)上的新PCT测定表明,即使在低浓度下也表现出优异的分析性能。观察到与Brahms Kryptor(R)相比轻微的偏差,但没有导致不适当的临床决策。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号