The design modification of advanced ventricular assist device to enhance pulse augmentation and regurgitant flow shut‐off

机译:先进心室辅助装置的设计改进,以增强脉冲增强和反弧流截止

获取原文
获取原文并翻译 | 示例
       

摘要

Abstract The new Advanced ventricular assist device (Advanced VAD) has many features such as improving pulsatility and preventing regurgitant flow during pump stoppage. The purpose of this study was to evaluate the effects of design modifications of the Advanced VAD on these features in vitro. Bench testing of four versions of the Advanced VAD was performed on a static or pulsatile mock loop with a pneumatic device. After pump performance was evaluated, each pump was run at 3000 rpm to evaluate pulse augmentation, then was stopped to assess regurgitant flow through the pump. There was no significant difference in pump performance between the pump models. The average pulse pressure in the pulsatile mock loop was 23.0, 34.0, 39.3, 33.8, and 37.3 mm Hg without pump, with AV010, AV020 3S, AV020 6S, and AV020 RC, respectively. The pulse augmentation factor was 48%, 71%, 47%, and 62% with AV010, AV020 3S, AV020 6S, and AV020 RC, respectively. In the pump stop test, regurgitant flow was ?0.60?±?0.70, ?0.13?±?0.57, ?0.14?±?0.09, and ?0.18?±?0.06 L/min in AV010, AV020 3S, AV020 6S, and AV020 RC, respectively. In conclusion, by modifying the design of the Advanced VAD, we successfully showed the improved pulsatility augmentation and regurgitant flow shut‐off features.
机译:摘要新的先进心室辅助装置(高级VAD)具有许多功能,如提高脉动性和防止泵停止期间的反尿剂流动。本研究的目的是评估先进的VAD在体外这些特征的设计修饰的影响。在具有气动装置的静态或脉动模拟环上执行四种版本的高级VAD的台阶测试。在评估泵性能后,每个泵以3000rpm运行以评估脉冲增强,然后停止以评估通过泵的反流流。泵模型之间的泵性能没有显着差异。脉冲模拟回路中的平均脉冲压力为23.0,34.0,39.3,33.8和37.3mm Hg,无泵,AV010,AV020 3S,AV020 6S和AV020 RC。脉冲增强因子分别为AV010,AV020 3S,AV020 6S和AV020 RC为48%,71%,47%和62%。在泵停止试验中,渐变流动是α0.60?±0.70,?0.13?±0.57,?0.14?±0.09,α0.18?0.18?0.18?0.18?0.18?0.18?010,AV020 3S,AV020 6S,以及AV020 RC分别。总之,通过修改先进VAD的设计,我们成功地展示了改善的脉动增强和反鲁的流动关闭功能。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号