首页> 外文期刊>Applied biosafety: Journal of the American Biological Safety Association. >Multicycle Autoclave Decontamination of N95 Filtering Facepiece Respirators
【24h】

Multicycle Autoclave Decontamination of N95 Filtering Facepiece Respirators

机译:N95过滤面板呼吸器的多网式高压灭弧室

获取原文
获取原文并翻译 | 示例
           

摘要

Introduction: During pandemic situations like the one caused by the emergent coronavirus SARS-CoV-2, healthcare systems face the challenge of limited personal protective equipment and impaired supply chains. This problem poses a threat to healthcare workers, first responders, and the public, which demands solutions that can span the gap between institutional shortages and resupplies. Objectives: To examine the efficacy of autoclave-based decontamination for the reuse of single-use surgical masks and N95 filtering facepiece respirators (FFRs). This method is the most readily available form of decontamination in the hospital and laboratory settings. Methods: Three models of N95 FFRs and two procedural masks were evaluated in this study. A moist heat autoclave using four different autoclave cycles: 115°C for one hour, 121.1°C for 30 minutes, 130°C for two minutes, and 130°C for four minutes was used. After the autoclave process, the FFRs were NIOSH fit tested and particle counting was performed for both coarse particles of 5 micrometers (μM) and fine particles from 0.1μM to 1.0μM. Results: We observed negligible alterations in the functionality and integrity of 3M 1805 and 3M 1870/1870+ N95 FFRs after three autoclave cycles. Surgical masks also showed minimal changes in functionality and integrity. The 3M 1860 FFR failed fit test after a single autoclave decontamination cycle. Discussion and Conclusion: The study finds that specific surgical masks and N95 FFR models can withstand autoclave decontamination for up to three cycles. Additionally, the autoclave cycles tested were those that could be readily achieved by both clinical and research institutions.
机译:介绍:在大流行的情况下,如由新兴冠状病毒SARS-COV-2引起的,医疗系统面临有限的个人防护设备和供应链受损的挑战。这个问题对医疗工作者,第一个受访者和公众构成了威胁,这需要解决可能跨越机构短缺和重新扫描之间的差距的解决方案。 目的:检查高压釜的去污的功效,用于重用一次性手术面罩和N95过滤面板呼吸器(FFR)。该方法是医院和实验室环境中最容易获得的净化形式。 方法:在本研究中评估了三种型号的N95 FFR和两个程序面具。使用四种不同的高压釜循环的湿热高压釜:115℃持续1小时,121.1℃,30分钟,130℃,使用130℃,4分钟。在高压釜过程之后,FFR是Niosh配合测试,并对粗颗粒为5微米(μm)和0.1μm至1.0μm的细颗粒进行颗粒计数。 结果:在三个高压灭菌循环后,我们观察到3M 1805和3M 1870/1870 + N95 FFR的功能和完整性的可忽略不计。外科口罩还表现出功能性和完整性的最小变化。在单个高压灭弧室循环循环后,3M 1860 FFR失效。 讨论和结论:该研究发现特定的外科助手和N95 FFR模型可以承受高达三个循环的高压灭菌。另外,测试的高压釜循环是可以通过临床和研究机构容易地实现的高压釜循环。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号