Biomonitoring method for the analysis of chromium and cobalt in human whole blood using inductively coupled plasma-kinetic energy discrimination-mass spectrometry (ICP-KED-MS)

机译:使用电感耦合等离子体 - 动能辨别 - 质谱法(ICP-KED-MS)使用电感耦合等血液中铬和钴的生物监测方法

获取原文
获取原文并翻译 | 示例
           

摘要

The Centers for Disease Control and Prevention developed a biomonitoringmethod to rapidly and accurately quantify chromium and cobalt in human whole blood by ICP-MS. Many metal-on-metal hip implants which contain significant amounts of chromium and cobalt are susceptible to metal degradation. This method is used to gather population data about chromium and cobalt exposure of the U.S. population that does not include people that have metal-on-metal hip implants so that reference value can be established for a baseline level in blood. We evaluated parameters such as; helium gas flow rate, choice and composition of the diluent solution for sample preparation, and sample rinse time to determine the optimal conditions for analysis. The limits of detection for chromium and cobalt in blood were determined to be 0.41 and 0.06 mu g L-1, respectively. Method precision, accuracy, and recovery for this method were determined using quality control material created in-house and historical proficiency testing samples. We conducted experiments to determine if quantitative changes in the method parameters affect the results obtained by changing four parameters while analyzing human whole blood spiked with National Institute of Standard and Technology traceable materials: the dilution factor used during sample preparation, sample rinse time, diluent composition, and kinetic energy discrimination gas flow rate. The results at the increased and decreased levels for each parameter were statistically compared to the results obtained at the optimized parameters. We assessed the degree of reproducibility obtained under a variety of conditions and evaluated the method's robustness by analyzing the same set of proficiency testing samples by different analysts, on different instruments, with different reagents, and on different days. The short-term stability of chromium and cobalt in human blood samples stored at room temperature was monitored over a time period of 64 hours by diluting and analyzing samples at different time intervals. The stability of chromium and cobalt post-dilution was also evaluated over a period of 48 hours and at two storage temperatures (room temperature and refrigerated at 4 degrees C). The results obtained during the stability studies showed that chromium and cobalt are stable in human blood for a period of 64 hours.
机译:疾病控制和预防的中心开发了一种生物监测方法,通过ICP-MS迅速准确地量化人类全血中的铬和钴。许多含有大量铬和钴的金属髋关节植入物易受金属降解的影响。该方法用于收集美国群体的铬和钴暴露的人口数据,这些方法不包括具有金属型髋关节植入物的人,以便可以为血液中的基线水平建立参考值。我们评估了如下参数;氦气流速,选择和组成样品制备的稀释剂溶液,并样品漂洗时间来确定分析的最佳条件。测定血液中铬和钴的检测限为0.41和0.06μg1-1。使用内部内部和历史熟练程度测试样品的质量控制材料确定该方法的方法精度,精度和恢复。我们进行了实验,以确定方法参数的定量变化是否会影响通过改变四个参数而获得的结果,同时分析了与国家标准和技术可追踪材料研究所的人类全血:样品制备过程中使用的稀释因子,样品漂洗时间,稀释剂组合物和动能辨别气体流速。与在优化参数上获得的结果相比,每个参数的增加和降低的结果与在优化参数上获得的结果进行了统计学。我们评估了在各种条件下获得的可重复性程度,并通过在不同的分析师在不同的仪器上分析了不同的分析师,不同的试剂以及不同的日子来评估该方法的鲁棒性。通过在不同时间间隔稀释和分析样品,在64小时内监测在室温下储存在室温下的人血液样品中的铬和钴的短期稳定性。还在48小时内和两个储存温度(室温和4℃下冷冻)的时间内评估铬铝醇和钴的稳定性。在稳定性研究期间获得的结果表明,铬和钴在人血中稳定为64小时。

著录项

  • 来源
    《Analytical methods》 |2017年第23期|共13页
  • 作者单位

    Ctr Dis Control &

    Prevent Inorgan &

    Radiat Analyt Toxicol Branch Natl Ctr Environm Hlth 4770 Buford Highway Mailstop F-50 Atlanta GA 30341 USA;

    Ctr Dis Control &

    Prevent Inorgan &

    Radiat Analyt Toxicol Branch Natl Ctr Environm Hlth 4770 Buford Highway Mailstop F-50 Atlanta GA 30341 USA;

    Ctr Dis Control &

    Prevent Inorgan &

    Radiat Analyt Toxicol Branch Natl Ctr Environm Hlth 4770 Buford Highway Mailstop F-50 Atlanta GA 30341 USA;

    Ctr Dis Control &

    Prevent Inorgan &

    Radiat Analyt Toxicol Branch Natl Ctr Environm Hlth 4770 Buford Highway Mailstop F-50 Atlanta GA 30341 USA;

    Ctr Dis Control &

    Prevent Inorgan &

    Radiat Analyt Toxicol Branch Natl Ctr Environm Hlth 4770 Buford Highway Mailstop F-50 Atlanta GA 30341 USA;

    Ctr Dis Control &

    Prevent Inorgan &

    Radiat Analyt Toxicol Branch Natl Ctr Environm Hlth 4770 Buford Highway Mailstop F-50 Atlanta GA 30341 USA;

  • 收录信息
  • 原文格式 PDF
  • 正文语种 eng
  • 中图分类 分析化学;
  • 关键词

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号