首页> 外文期刊>Anaesthesia: Journal of the Association of Anaesthetists of Great Britain and Ireland >The effect of rate of injection on injection pressure profiles measured using in‐line and needle‐tip sensors: an in‐vitro study
【24h】

The effect of rate of injection on injection pressure profiles measured using in‐line and needle‐tip sensors: an in‐vitro study

机译:注射率对使用在线和针尖传感器测量的注射压力分布的影响:体外研究

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

Summary Alongside ultrasonic visualisation, measurement of injection pressure is an effective tool for reducing the risk of intraneural injection during peripheral nerve block. The aim of this study was to compare injection pressure profiles when measured along the injection line with the pressure measured directly at the needle tip using different rates of injection. A syringe pump delivered a 5‐ml injection of saline into silicone gel at three different speeds (5?ml.min ?1 , 10?ml.min ?1 and 15?ml.min ?1 ). Fibreoptic pressure sensors recorded real‐time pressure profiles of the injection pressure directly at the needle tip and along the injection line. A total of 15 injections were successfully performed, five for each injection rate, totalling 30 recorded pressure profiles. More rapid rates of injection caused peak pressure measured in‐line to increase, whereas pressure measured at the needle tip remained constant (mean ( SD ) pressure in‐line 30.76 (3.45) kP a vs. 72.25 (1.55) kP a and mean ( SD ) pressure at needle tip 19.92 (1.22) kP a vs. 20.93 (2.66) kP a at 5?ml.min ?1 and 15?ml.min ?1 , respectively). Injection pressure profiles showed that in‐line pressure measurement failed to record precise real‐time pressure changes occurring at the needle tip (mean (95% CI ) pressure difference 10.8 (6.98–14.70) kP a vs. 51.2 (47.52–54.89) kP a for in‐line and needle‐tip measures, respectively). We conclude that, in order to accurately monitor the true injection pressure generated, independent from operator and injection parameters, measurement at the needle tip is necessary, as injection pressure measured along the injection line is an unreliable surrogate.
机译:发明内容与超声波可视化一起,注射压力的测量是一种有效的工具,用于降低周边神经块期间的内部注射风险。该研究的目的是在沿喷射管线测量时比较注射压力型材,其使用不同的注射率直接在针尖处测量的压力。注射器泵以三种不同的速度将5ml注入盐水注入硅胶凝胶中(5?ml.min?1,10?ml.min?1和15?ml.min?1)。纤维电压传感器直接在针尖和喷射管线处录制注射压力的实时压力轮廓。每次注射速率成功地进行了总共15个注射,总计30个记录的压力曲线。更快的注射率引起峰值压力在线测量以增加,而在针尖处测量的压力保持恒定(平均(SD)压力在线30.76(3.45)KP A与272.25(1.55)KP A和平均值( SD)针尖19.92(1.22)Kp A与20.93(2.66)Kp A分别为5?ml.minα1和15?ml.min?1。注射压力曲线显示,在线压力测量未能在针尖发生精确的实时压力变化(平均值(95%CI)压力差10.8(6.98-14.70)KP A和51.2(47.52-54.89)KP A分别用于直线和针尖测量。我们得出结论,为了准确地监测所产生的真正注射压力,独立于操作者和注射参数,需要在针尖处的测量是必要的,因为沿喷射管线测量的喷射压力是不可靠的替代。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号