首页> 外文期刊>Acta Radiologica >Iterative scatter correction for grid-less skeletal radiography allows improved image quality equal to an antiscatter grid in adjunct with dose reduction: a visual grading study of 20 body donors
【24h】

Iterative scatter correction for grid-less skeletal radiography allows improved image quality equal to an antiscatter grid in adjunct with dose reduction: a visual grading study of 20 body donors

机译:较少骨骼骨展放射线照相的迭代散射校正允许改善的图像质量等于具有剂量减少的辅助反射网格的图像质量:20个车身供体的视觉分级研究

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

Background Iterative scatter correction (ISC) is a new technique applicable to plain radiography; comparable to iterative reconstruction for computed tomography, it promises dose reduction and image quality improvement. ISC for bedside chest X-rays has been applied and evaluated for some time and has recently been commercially offered for plain skeletal radiography. Purpose To analyze the potential of ISC for plain skeletal radiography with regard to image quality improvement, dose reduction, and replacement for an antiscatter grid. Material and Methods A total of 385 radiographs with different imaging protocols of the pelvis and cervical spine were acquired from 20 body donors. Radiographs were rated by four radiologists. Ratings were analyzed with visual grading characteristics (VGC) analysis. The area under the VGC curve was used as a measure of difference in image quality. Results Without ISC, the grid-less images were rated significantly worse than their grid-based counterparts (0.389, P = 0.005); adding ISC made image quality equal (0.498; P = 0.963). In grid-less imaging, reduction of dose by 50% led to significant image quality impairment (0.415, P = 0.001); this was fully counterbalanced when ISC was added (0.512; P = 0.588). Conclusion ISC for plain skeletal radiography has the ability to replace the antiscatter grid without image quality impairment, to improve image quality in grid-less imaging, and to reduce patient radiation dose by 50% without substantial loss in image quality.
机译:背景技术迭代散射校正(ISC)是一种适用于普通射线照相的新技术;可与计算断层扫描的迭代重建相当,它承诺减少剂量和图像质量改进。床侧胸部X射线的ISC已被应用和评估一段时间,最近已为平原骨骼射线照相造影商业提供。目的分析ISC的潜力对于普通骨架射线照相,关于图像质量改善,剂量减少,更换防火电网。材料和方法总共385个具有不同成像协议的骨盆和颈椎的射线照片是从20个锻炼捐助者获得的。射线照片由四名放射科医师评定。通过视觉分级特性(VGC)分析分析评级。 VGC曲线下方的区域被用作图像质量差异的量度。没有ISC的结果,额外的网格图像比其基于网格的对应物(0.389,P = 0.005)显着差;添加ISC制作的图像质量等于(0.498; p = 0.963)。在较少的成像中,减少剂量50%导致显着的图像质量损伤(0.415,P = 0.001);当添加ISC时,这完全抵消(0.512; P = 0.588)。结论普通骨架射线照相的ISC能够更换不含图像质量损伤的防脆网格,以提高网格成像中的图像质量,并减少50%的患者辐射剂量,在图像质量中没有大量损失。

著录项

相似文献

  • 外文文献
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号