首页> 外文期刊>Acta Obstetricia et Gynecologica Scandinavica: Official Publication of the Nordisk Forening for Obstetrik och Gynekologi >Reproducibility and acceptability of ultrasound measurements of head–perineum distance
【24h】

Reproducibility and acceptability of ultrasound measurements of head–perineum distance

机译:头部阴部距离超声测量的再现性和可接受性

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

Abstract Introduction We aimed to test the reproducibility of head–perineum distance ( HPD ) measurements using two different ultrasound devices and five examiners, to compare ultrasound measurements and clinical assessments and to study if ultrasound examinations were acceptable for women in labor. Material and methods A reproducibility study was performed at Lund University Hospital, Sweden and Landspitali University Hospital, Iceland from February 2015 to February 2017. The study population comprised 40 healthy women in labor. HPD was measured with three replicate measurements from each woman with two different ultrasound devices, and the measurements were compared with clinical assessments. Acceptability was tested with a visual analog scale ( VAS ), and the mean VAS score from both ultrasound devices was compared with the VAS score from clinical palpation. Results The median time interval between start of examinations with devices was 10 min (range 1–26 min). The intra‐observer repeatability coefficient was 4.3 mm and the intraclass correlation coefficient was 0.97 (95% CI 0.95–0.98). The intraclass correlation coefficient between the two devices was 0.86 (95% CI 0.74–0.93) and limits of agreement were ?9.6 mm to 16.6 mm. However, we observed a significant mean HPD difference between devices (3.5 mm; 95% CI 1.4–5.6 mm). Clinical assessments and the mean measurements of HPD were correlated ( r = 0.64, p 0.01). We found significant differences in acceptability in favor of ultrasound. The mean VAS score for both ultrasound devices was 2.0 vs. 4.1 for clinical examination ( p 0.01). Conclusion We found excellent intra‐observer repeatability, good correlation but significant difference between devices. Women reported less discomfort with ultrasound than with clinical examinations.
机译:摘要介绍旨在使用两种不同的超声装置和五次审查员测试头部阴部距离(HPD)测量的再现性,以比较超声测量和临床评估,并在劳动中妇女接受超声检查。材料和方法在2015年2月至2017年2月冰岛德蒙登大学医院,瑞典和地地性大学医院进行再现性研究。该研究人口占劳动力40名健康妇女。用来自两个不同超声装置的每个妇女的三次复制测量测量HPD,并将测量与临床评估进行比较。使用视觉模拟量表(VAS)测试可接受性,并将来自临床触诊的VAS分数进行比较来自两个超声装置的平均VAS评分。结果使用器件开始的检查开始之间的中值时间间隔为10分钟(范围1-26分钟)。观察者内的重复性系数为4.3mm,腹部相关系数为0.97(95%CI 0.95-0.98)。两种器件之间的腹部相关系数为0.86(95%CI 0.74-0.93),协议的限制为9.6毫米至16.6毫米。然而,我们观察到设备之间的显着平均HPD差异(3.5毫米; 95%CI 1.4-5.6mm)。临床评估和HPD的平均测量是相关的(R = 0.64,P <0.01)。我们发现有利于超声波的可接受性的显着差异。对于临床检查,两个超声装置的平均VAS分数为2.0与4.1(P <0.01)。结论我们发现了出色的观察室内的重复性,相关性良好但设备之间的显着差异。妇女报告的超声波不适于临床检查。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号