首页> 外文期刊>Cryogenics >Low-frequency magnetic Interference in a SQUID gradiometer caused by a Cryotiger~R gas-mixture cooler
【24h】

Low-frequency magnetic Interference in a SQUID gradiometer caused by a Cryotiger~R gas-mixture cooler

机译:Cryotiger〜R混合气体冷却器在SQUID梯度计中产生低频电磁干扰

获取原文
获取原文并翻译 | 示例
获取外文期刊封面目录资料

摘要

A Cryotiger~R gas-mixture cooler was applied for cooling of three high-Tc SQUID magnetometers. These SQUID magnetometers were mounted on an alumina holder in an axial gradiometer configuration. From 20 Hz upward, the system noise was about 0.1 pT/ V/Rz. Below this frequency, the noise gradually increased to a level of 10 pT/Hz at 1 Hz. This low-frequency excess noise appeared to be due to remnant magnetization of the Cryotiger cold head. Movement of magnetic cold-head parts with respect to the SQUIDs are induced by pressure fluctuations in the heat exchanger lines. By using one SQUID as a reference for the cooler noise, a first-order gradiometer can be formed in which the cooler noise is eliminated. To establish a proper second-order gradiometer either a fourth SQUID has to be added, or the spatial separation between cold head and SQUIDs has to be increased significantly.
机译:将Cryogerger R气体混合物冷却器用于冷却三个高Tc SQUID磁力计。这些SQUID磁力计以轴向梯度计配置安装在氧化铝支架上。从20 Hz开始,系统噪声约为0.1 pT / V / Rz。低于此频率,噪声在1 Hz时逐渐增加到10 pT / Hz的水平。低频过量噪声似乎是由于低温冷冻头的剩余磁化引起的。磁性冷头部件相对于SQUID的运动是由热交换器管路中的压力波动引起的。通过使用一个SQUID作为冷却噪声的参考,可以形成一阶梯度仪,其中消除了冷却噪声。为了建立合适的二阶梯度仪,必须添加第四个SQUID,或者必须大幅增加冷头和SQUID之间的空间间隔。