首页> 外文期刊>Cryogenics >Construction and test of a cryocooler-cooled low-Tc SQUID gradiometer system
【24h】

Construction and test of a cryocooler-cooled low-Tc SQUID gradiometer system

机译:冷却器冷却的低Tc SQUID梯度仪系统的构建和测试

获取原文
获取原文并翻译 | 示例

摘要

A cryocooler-cooled low-Tc SQUID gradiometer system was designed and constructed. It is based on a 4 K GM-cooler that is positioned several meters from the measuring head containing the SQUID sensors. The cooling power is transported by means of a circulating helium flow. Thus, there are three main components: the cooler vessel, the measuring head and the connecting flow-line. In stable operation, the SQUID and gradiometric pick-up loop are about 4.3 K. The overall heat loads to the cooler 1st and 2nd stages are estimated at 17 W and 0.3 W, respectively. The overall system noise was about 10 fT/√Hz although a number of peaks were observed in the spectrum. As an example, an adult heart beat was recorded.
机译:设计并构建了一个冷却器冷却的低Tc SQUID梯度仪系统。它基于一个4 K GM冷却器,该冷却器距离包含SQUID传感器的测量头几米远。冷却功率通过循环氦流传输。因此,有三个主要组成部分:冷却容器,测量头和连接流水线。在稳定运行中,SQUID和梯度拾取回路约为4.3K。估计冷却器第一级和第二级的总热负荷分别为17 W和0.3W。尽管在频谱中观察到许多峰值,但整个系统的噪声约为10 fT /√Hz。例如,记录了成人心跳。