We reviewed 51 plate fixations for fracture of the humeral diaphysis managed during the last 3 years. The indications for surgery included non-union, delayed union, polytrauma, neurological involvement and inadequate reduction. All the fractures were stabilized with a Limited Contact Dynamic Compression Plate (LCDC Plate). We found that this plate gave adequate rotational and vertical stability to the fracture. 48 fractures healed after median 2.5 months. Early evidence of callus formation indicated minimal compromise with the vascularity of bone which may be attributed to the undercut design of the plate, causing relatively less cortical ischemia. In addition, the plate fixation was associated with fewer complications, as compared to those reported with intramedullary nailing.
机译:我们回顾了最近3年内处理的51块钢板固定物治疗的肱骨干physi端骨折。手术适应症包括不愈合,延迟愈合,多发伤,神经系统受累和复位不足。所有骨折均用有限接触动态压缩板(LCDC板)稳定。我们发现该板为骨折提供了足够的旋转和垂直稳定性。中位2.5个月后,有48处骨折he愈。愈伤组织形成的早期证据表明与骨血管的损害最小,这可能归因于板的底切设计,从而导致相对较少的皮质缺血。此外,与髓内钉术相比,钢板固定的并发症更少。