Minimisation of pressure dependent mass discrimination in the ion source of the Helix MC Plus noble gas mass spectrometer

机译:最小化螺旋MC的离子源中的压力依赖质量辨别:斜体>加上惰性气体质谱仪

获取原文
获取原文并翻译 | 示例
获取外文期刊封面目录资料

摘要

AbstractWhen tuned for maximum sensitivity, the electron bombardment ion source on the ANU (the Australian National University) Helix MCPlusnoble gas mass spectrometer produces an Ar ion beam that is mass discriminated as a function of Ar partial pressure. This effect can be reduced substantially or eliminated by adjusting the repeller and trap voltages, although with a sensitivity loss of up to 50%. A recommended procedure for tuning an ion source to reduce pressure dependent mass discrimination is described. Similar procedures can be applied to other noble gases and other ion sources to ensure that pressure dependent mass discrimination is avoided.]]>