Simultaneous multi-site measurement system for the assessment of pulse wave delays

机译:同时多站点测量系统,用于评估脉搏波延迟

获取原文
获取原文并翻译 | 示例

摘要

A precise, multi-track system for the simultaneous, real-time measurement of electrocardiographic (ECG) and many photopletysmographic (PPG) signals is described. This system allows the calculation of pulse wave delay parameters such as pulse arrival time (PAT) and pulse transit time (PTT). The measurement system was built on a custom, real-time embedded system with multiple specific analogue-front-end devices. Signals were recorded on-line and data were processed off-line in the Matlab software. Testing of human subjects was carried out on a group of 16 volunteers. The system was capable of taking a measurement of one 24-bit ECG and eight 22-bit PPG tracks with high precision (input-referred noise 1.4 mu V for ECG and about 20 pA for PPG). All signals are sampled simultaneously (phase shift between ECG and PPG is only 1.5 ms for 250 Hz frequency sampling). Significant differences in pulse wave delays were found for the 16 subjects studied (e.g. about 100 ms for PAT on a right toe, 40 ms for differential PAT on left-right toes and about 100 ms for PTT calculated for forehead-right toe pulse wave). The proposed system provides a simultaneous and continuous evaluation of pulse wave delays for the entire arterial bed. The proposed measurement methods are comfortable and can be used for a long time. Simultaneous measurements of pulse wave delays at various sites increase the reliability of measurement and create new possibilities for medical diagnosis. (C) 2019 Nalecz Institute of Biocybernetics and Biomedical Engineering of the Polish Academy of Sciences. Published by Elsevier B.V. All rights reserved.
机译:描述了用于同时,实时测量心电图(ECG)和许多光照学(PPG)信号的精确多轨系统。该系统允许计算脉冲波延迟参数,例如脉冲到达时间(PAT)和脉冲传输时间(PTT)。测量系统是在具有多个特定模拟 - 前端设备的自定义实时嵌入式系统上构建的。在线路上记录信号,在Matlab软件中离线处理数据。人类受试者的测试是在一组16名志愿者进行的。该系统能够测量具有高精度的24位ECG和八个22位PPG轨道(用于ECG的输入引用噪声1.4μV以及PPG约20Pa)。所有信号同时采样(ECG和PPG之间的相移仅为250 Hz频率采样的1.5ms)。发现了用于研究的16个受试者的脉搏波延迟的显着差异(例如,在右脚趾上拍拍的约100ms,对于左右脚趾的差异拍拍40 ms,对于针对前额右脚趾脉冲波的PTT约为100 ms。 。所提出的系统提供了对整个动脉床的脉搏波延迟的同时和连续评估。所提出的测量方法舒适,可长时间使用。各个站点的脉冲波延迟的同时测量增加了测量的可靠性,并为医学诊断产生了新的可能性。 (c)2019年纳雷斯州博士科学学院生物医学研究所。 elsevier b.v出版。保留所有权利。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号