Standardization of assay methods reduces variability of total PSA measurements: An Irish study

机译:测定方法的标准化降低了PSA测量总量的可变性:爱尔兰研究

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

Study Type - Diagnosis (quality control) Level of Evidence 2b What's known on the subject? and What does the study add? Today, numerous assays for PSA detection are available from various manufacturers. However, these various assays do not detect PSA equally and several studies have demonstrated variability between them. In order to harmonise PSA results and reduce the discrepancies, reference materials are available for assay calibration. We have demonstrated significantly variability between 6 different assay methods currently in use in 9 hospitals despite assay calibration. Variability in PSA values was reduced with the standardisation of the assay method in 4 hospitals. Our results highlight the dilemma of PSA assay variability and stress the need for nationwide standardisation of PSA testing. Objective To determine whether standardization of total prostate-specific antigen (tPSA) assay methods reduces variability in tPSA measurements. Patients and methods Blood samples from 84 patients attending a single urology department were distributed across nine hospitals selected throughout Ireland for the independent determination of tPSA under the same conditions. The selected hospitals collectively used six different assay methods for tPSA detection: Beckman Hybritech WHO Calibrated (used as reference method), Tosoh AIA 1800, Roche E170 (used in three hospitals), Abbott AxSYM, Immulite 2500 2nd Generation (used in two hospitals) and Siemens ADVIA Centaur. The method of tPSA detection was next standardized in a subset of four hospitals using the same assay method and the measurements were repeated. The difference in mean tPSA in the cohort across the hospitals tested was determined and the Bland-Altman test was used to assess the agreement between each test. Analysis was performed over both the full (0.5-30 μg/L, N= 84) and a narrow (3-7 μg/L, n= 25) tPSA range. Results The range and the mean tPSA of the full cohort were inflated across the eight test hospitals, when compared with the reference hospital. The poorest agreement between assay methods was associated with a bias of 2.2 ± 2.4 μg/L. The variability in tPSA measurements between assay methods was inconsistent across the range of tPSA values tested and increased with increasing mean tPSA. Agreement in reported tPSA was excellent after standardization of tPSA assay methods (bias <0.2 μg/L). Over the narrow 3-7 μg/L PSA range, 12/25 (48%) patients had a tPSA range of values across all hospitals in excess of 2 μg/L. Following standardization of the tPSA assay method, patient tPSA ranges were <0.5 μg/L for 13/25 (52%) patients. Conclusions We have shown that the lack of standardization of tPSA assay methods across a panel of Irish hospitals leads to significant variability in the measured tPSA values for the same patient samples. Variability in tPSA values was reduced with the standardization of the assay method in four hospitals. Standardization of PSA testing on a nationwide scale is warranted.
机译:研究类型 - 诊断(质量控制)证据级别2B对该主题所知的内容?这项研究添加了什么?如今,各种制造商可获得许多PSA检测的测定。然而,这些各种测定不能同样检测PSA,并且有几项研究表明它们之间的可变性。为了协调PSA结果并降低差异,可用于测定校准的参考资料。我们已经在9家医院使用的6种不同的测定方法之间证明了6种不同的测定方法之间的显着变化。通过4家医院的测定方法的标准化降低了PSA值的可变性。我们的结果突出了PSA测定变异性的困境和对PSA测试的全国范围标准化的需求。目的确定总前列腺特异性抗原(TPSA)测定方法的标准化是否降低了TPSA测量的可变性。患者和方法来自84名参加单一泌尿外科患者的血液样本分布在整个爱尔兰选择的九个医院,以在同一条件下独立测定TPSA。所选医院共同使用六种不同的测定方法,用于TPSA检测:Beckman Hybritech校准(用作参考方法),Tosoh Aia 1800,Roche E170(用于三家医院),Abbott Axsym,Immulite 2500第2代(在两家医院使用)和西门子Advia Centaur。接下来,TPSA检测方法在使用相同的测定方法的四个医院的子集中标准化,重复测量。确定测试的队列中的群组中的平均TPSA差异,并使用Bland-Altman测试来评估每个测试之间的协议。通过完全(0.5-30μg/ L,n = 84)和窄(3-7μg/ L,n = 25)TPSA范围进行分析。结果与参考医院相比,八个测试医院的整个队列的范围和平均TPSA膨胀。测定方法之间最糟糕的一致性与2.2±2.4μg/ L的偏差有关。测定方法之间的TPSA测量的可变性在测试的TPSA值范围内不一致,随着平均TPSA的增加而增加。报告的TPSA协议在TPSA测定方法的标准化后具有优异的(偏置<0.2μg/ L)。在狭窄的3-7μg/ l psa范围内,12/25(48%)患者在所有医院的TPSA范围内有超过2μg/升的医院。在TPSA测定方法的标准化之后,患者TPSA范围为13/25(52%)患者<0.5μg/ L.结论我们已经表明,在爱尔兰医院面板上缺乏TPSA测定方法的标准化导致相同患者样品的测量TPSA值中的显着变化。通过四家医院的测定方法的标准化减少了TPSA值的可变性。保证了全国范围的PSA测试的标准化。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号