Determination of UV filters in high ionic strength sample solutions using matrix-compatible coatings for solid-phase microextraction

机译:用基质兼容涂层测定高离子强度样品溶液中的UV过滤器,用于固相微萃取

获取原文
获取原文并翻译 | 示例
           

摘要

A double-confined polymeric ionic liquid (PIL) sorbent coating was fabricated for the determination of nine ultraviolet (UV) filters in sample solutions containing high salt content by direct immersion solid-phase microextraction (DI-SPME) coupled to high-performance liquid chromatography (HPLC). The IL monomer and crosslinker cations and anions, namely, 1-vinyl-3-decylimidazolium styrenesulfonate ([VImC(10)] [SS]) and 1,12-di (3-vinylbenzylimidazolium) dodecane distyrenesulfonate ([(VBIm)(2)C-12] 2 [SS]), were co-polymerized to create a highly stable sorbent coating which allowed for up to 120 direct-immersion extractions in 25% NaCl (w/v) solution without a decrease in its extraction capability. Extraction and desorption parameters such as desorption solvent, agitation rate, extraction time, desorption solvent volume, and desorption time were evaluated and optimized. The analytical performance of the styrenesulfonate anion-based PIL fiber, PIL fiber containing chloride anions, and a commercially available polydimethylsiloxane/divinylbenzene (PDMS/DVB) fiber were compared. Coefficients of determination (R-2) for the styrenesulfonate anion-based PIL fiber ranged from 0.995 to 0.999 and the limits of detection (LODs) varied from 0.1 to 5 mu g L-1. The developed method was successfully applied in real water samples including tap, pool, and lake water, and acceptable relative recovery values were obtained. The lifetime of the PIL fiber containing chloride anions as well as the PDMS/DVB fiber were considerably shorter than the PIL fiber containing the styrenesulfonate anion, with both fibers showing a notable decrease in reproducibility and significant damage to the sorbent coating surface after 40 and 70 extractions, respectively. The R-2 values for the chloride anion containing PIL fiber were at or higher than 0.991 with LODs ranging from 0.5 to 5 mu g L-1. For the PDMS/DVB fiber, R-2 values ranged from 0.992 to 0.999 and LODs were found to be as low as 0.2 mu g L-1 a
机译:一种双局限于聚合离子液体(PIL)吸附剂涂层被制造为9对紫外线(UV)在含有由耦合到高效液相色谱法直接浸入固相微萃取(DI-SPME)高盐含量的样品溶液过滤器的判定(HPLC)。所述IL单体和交联剂的阳离子和阴离子,即,1-乙烯基-3- decylimidazolium苯乙烯([VIMC(10)] [SS])和1,12-二(3- vinylbenzylimidazolium)十二烷distyrenesulfonate([(VBIm)(2 )C-12] 2 [SS]),进行共聚合,以创建其允许多达120直接浸没萃取在25%NaCl的高度稳定的吸附剂涂层(W / v),而不在其萃取能力的降低的解决方案。萃取和解吸参数,如溶剂解吸,搅拌速率,提取时间,解吸溶剂体积,和脱附时间进行了评价和优化。所述基于苯乙烯阴离子PIL纤维的分析性能,含有氯化物阴离子,和可商购的聚二甲基硅氧烷/二乙烯基苯PIL纤维(PDMS / DVB)纤维进行了比较。对于基于阴离子苯乙烯PIL纤维测定的系数(R-2)的范围为0.995至0.999和检测(检测限)的极限为0.1〜5微米克L-1变化。所开发的方法实际水样,包括抽头,游泳池和湖水,和可接受的相对回收率值获得在已成功应用。含有氯阴离子以及所述PDMS的PIL纤维的寿命/ DVB纤维比含有苯乙烯磺酸盐阴离子的PIL纤维短得多,具有两种纤维表示再现性的显着降低和显著损坏吸附剂涂覆表面40和70后提取,分别。为氯阴离子含有PIL纤维的R-2值是在等于或高于0.991与检测限为0.5至5微米克L-1。对于PDMS / DVB纤维,R-2的值的范围为0.992到0.999,并且发现检测限是低至0.2亩克L-1中的

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号