TomoTherapy treatments of multiple brain lesions: an in-phantom accuracy evaluation

机译:多发性脑病变的清汤治疗:幻影精度评估

获取原文
获取原文并翻译 | 示例

摘要

Aim of the present study was to evaluate the accuracy which can be obtained with helical TomoTherapy (R) (HT, Accuray) systems in the case of multiple intracranial targets treatments. Set-up accuracy was measured, for different registration options and MegaVoltage CT (MVCT) slice thickness, by applying known misalignments to an ad-hoc developed phantom. End-to-end (E2E) tests were performed to assess the delivery accuracy in phantoms containing multiple targets by using radiochromic films: measured dose distribution centroids were compared with physical and calculated target positions on axial and coronal planes. A Gamma index analysis was carried out on planned and measured planar dose maps. The bone and tissue algorithm with the fine MVCT reconstruction grid gave the best results among the automatic options. The most accurate registration modality resulted to be the manual one with a sub-voxel accuracy shifts and a capability in the detection of rotations within 0.3 degrees. For the E2E along the coronal plane (six targets), a mean deviation between measured dose distribution centroids and physical barycenters of 0.6 mm (range 0.1 mm-1.3 mm) was observed. Along the axial plane (five targets), a mean deviation of 1.2 mm (range 0.7 mm-2.1 mm) was found for the centroids shifts. Gamma index (5%, 1 mm, local) passing rates higher than 87.5% between planned and delivered dose distributions were measured. These results demonstrate that multiple brain lesion HT treatments are feasible with an accuracy at least comparable to frameless linac-based delivery, when a set-up capable to assure angular corrections and a reliable patient immobilization is employed.
机译:本研究的目的是评估在多个颅内靶标处理的情况下用螺旋疗法(R)(R)(R)(R)(R)(R)(R)(R)(R)(HT)(R)(HT)(HT)(R)(R)(R)(R)(R)(R)(R)(R)(R)(R)系统提供的精度。通过将已知的未对准应用于Ad-hoc开发的幻影,测量了设置精度,用于不同的登记选项和MegaVoltage CT(MVCT)切片厚度。进行端到端(E2E)测试以评估通过使用放射褐色膜的含有多个靶标的含有多种靶的含量的递送精度:将测量剂量分布质细胞与轴向和冠状平面上的物理和计算的目标位置进行比较。在计划和测量的平面剂量图上进行了伽玛指数分析。具有精细MVCT重建网格的骨骼和组织算法在自动选项中提供了最佳效果。最精确的注册模型导致手动具有子体素精度偏移的手动,并且在0.3度之内检测旋转的能力。对于沿着冠状平面(六个靶标)的E2E,观察到测量剂量分布质心和0.6mm(范围0.1mm-1.3mm)之间的平均偏差。沿轴向平面(五个靶),发现质心偏移的平均偏差为1.2mm(范围0.7mm-2.1mm)。测量伽马指数(5%,1毫米,局部)在计划和交付剂量分布之间的速率高于87.5%。这些结果表明,当能够确保角度校正和可靠的患者固定化的设置时,多种脑病变HT处理具有至少与基于无框架的LinaC的递送的准确性。

著录项

  • 来源
    《Physics in medicine and biology. 》 |2019年第2期| 共11页
  • 作者单位

    Dept Biomed Expt &

    Clin Sci Mario Serio Viale Morgagni 50 I-50134 Florence Italy;

    Azienda Osped Univ Careggi Med Phys Unit Largo Brambilla 3 I-50134 Florence Italy;

    Dept Biomed Expt &

    Clin Sci Mario Serio Viale Morgagni 50 I-50134 Florence Italy;

    Dept Biomed Expt &

    Clin Sci Mario Serio Viale Morgagni 50 I-50134 Florence Italy;

    Dept Biomed Expt &

    Clin Sci Mario Serio Viale Morgagni 50 I-50134 Florence Italy;

    Azienda Osped Univ Careggi Radiotherapy Unit Largo Brambilla 3 I-50134 Florence Italy;

    Dept Biomed Expt &

    Clin Sci Mario Serio Viale Morgagni 50 I-50134 Florence Italy;

    Dept Biomed Expt &

    Clin Sci Mario Serio Viale Morgagni 50 I-50134 Florence Italy;

    Dept Biomed Expt &

    Clin Sci Mario Serio Viale Morgagni 50 I-50134 Florence Italy;

  • 收录信息
  • 原文格式 PDF
  • 正文语种 eng
  • 中图分类 R35;
  • 关键词

    TomoTherapy; multiple brain metastasis; in phantom setup accuracy;

    机译:Tomotherapy;多脑转移;在幻影设置精度;

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号