首页> 外文期刊>Physics in medicine and biology. >Quality assurance of gating response times for surface guided motion management treatment delivery using an electronic portal imaging detector
【24h】

Quality assurance of gating response times for surface guided motion management treatment delivery using an electronic portal imaging detector

机译:使用电子门户成像检测器的表面导向运动管理处理递送的门控响应时间的质量保证

获取原文
获取原文并翻译 | 示例
           

摘要

Abstract, Gating response times of monitoring system/LINAC combinations for gated radiotherapy treatments are notoriously difficult to measure. This difficulty may reflect why the incorporation of gating response times is not thoroughly considered when creating treatment margins for gated radiotherapy treatments, however ignoring the effect of gating response time could lead to significant treatment inaccuracies. This study shows a methodology which measured gating response times for AlignRT/Varian Truebeam combination which appears to be applicable to any surface guided monitoring/Linac combination as well as those combinations which incorporate extended lag times (>200ms). Beam on lag time appears to be measurably greater than beam off lag time. Although these gating response times are slower than other gating systems on the market, the advantages surface guided radiotherapy (SGRT) could potentially provide for treatment accuracy is discussed as well as demonstrating the lack of guidance regarding SGRT with respect to gated treatments
机译:摘要,监测系统/ LINAC组合的门控响应次数难以测量。这种困难可能反映为什么在为门控放射治疗治疗产生治疗边距时不彻底地考虑纳入响应时间的原因,但是忽略了门控响应时间的影响可能导致显着的治疗不准确性。该研究表明了一种方法学,其测量了对准的响应时间,用于对准的响应时间,其似乎适用于任何表面引导监测/ LINAC组合以及包含延长滞后时间(> 200ms)的组合。滞后时间的光束似乎可测量地大于截止滞后时间。虽然这些门控响应时间比市场上的其他门控系统慢慢,但讨论了表面引导放射疗法(SGRT)的优点是讨论了治疗准确性,并展示了与门控处理的SGRT缺乏指导

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号