首页> 外文期刊>Journal of chromatography, A: Including electrophoresis and other separation methods >Application of L-cystine modified zeolite for preconcentration and determination of ultra-trace levels of cadmium by flame atomic absorption spectrometry
【24h】

Application of L-cystine modified zeolite for preconcentration and determination of ultra-trace levels of cadmium by flame atomic absorption spectrometry

机译:L-胱氨酸改性沸石在火焰原子吸收光谱法富集和测定痕量镉中的应用

获取原文
获取原文并翻译 | 示例
           

摘要

A very convenient, sensitive and precise solid phase extraction (SPE) system was developed for enrichment and determination of ultra-trace of cadmium ion in water and plant samples. This method was based on the retention of cadmium(II) ions by L-cystine adsorbed in Y-zeolite and carry out in a packed mini-column. The retained cadmium ions then were eluted and determined by flame atomic absorption spectrometry. The scanning electron microscopy (SEM), powder X-ray diffraction (XRD) and Fourier Transform Infrared (FT-IR) spectroscopy techniques were applied for the characterization of cystine modified zeolite (CMZ). Some experimental conditions affecting the analytical performance such as pH, eluent type, concentration of sample, eluent flow rate and also the presence of interfering ions were investigated. The calibration graph was linear within the range of 0.1-7.5 ng mL(-1) and limit of detection was obtained 0.04 ng mL(-1) with the preconcentration factor of 400. The relative standard deviation (RSD) was obtained 1.4%, indicating the excellent reproducibility of this method. The proposed method was successfully applied for the extraction and determination of cadmium(II) ion in black tea, cigarette's tobacco and also various water samples. (C) 2016 Elsevier B.V. All rights reserved.
机译:开发了一种非常方便,灵敏且精确的固相萃取(SPE)系统,用于富集和测定水和植物样品中的痕量镉离子。该方法基于吸附在Y沸石中的L-胱氨酸对镉(II)离子的保留,并且在填充的微型柱中进行。然后洗脱保留的镉离子,并通过火焰原子吸收光谱法测定。扫描电子显微镜(SEM),粉末X射线衍射(XRD)和傅里叶变换红外(FT-IR)光谱技术被用于表征胱氨酸修饰的沸石(CMZ)。研究了一些影响分析性能的实验条件,例如pH,洗脱液类型,样品浓度,洗脱液流速以及干扰离子的存在。校正曲线在0.1-7.5 ng mL(-1)范围内是线性的,检出限为0.04 ng mL(-1),预浓系数为400。相对标准偏差(RSD)为1.4%,表明该方法具有极好的重现性。该方法已成功用于红茶,卷烟,烟草和各种水样中镉(II)离子的提取和测定。 (C)2016 Elsevier B.V.保留所有权利。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号