首页> 外文期刊>The Journal of Urology >Re: Automated renal stone volume measurement by noncontrast computerized tomography is more reproducible than manual linear size measurement: S. R. Patel, P. Stanton, N. Zelinski, E. J. Borman, M. A. Pozniak, S. Y. Nakada and P. J. Pickhardt J Urol 2011; 186: 2275-2279
【24h】

Re: Automated renal stone volume measurement by noncontrast computerized tomography is more reproducible than manual linear size measurement: S. R. Patel, P. Stanton, N. Zelinski, E. J. Borman, M. A. Pozniak, S. Y. Nakada and P. J. Pickhardt J Urol 2011; 186: 2275-2279

机译:回复:通过非对比计算机断层摄影术自动测量肾结石体积比手动线性尺寸测量具有更高的重现性:S。R. Patel,P。Stanton,N。Zelinski,E。J. Borman,M。A. Pozniak,S。Y. Nakada和P. J. Pickhardt J Urol 2011; 186:2275-2279

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

To the Editor: In their discussion Patel et al note that the accuracy of automated volume measurements was verified in other models using phantoms. However, they were using coronary artery calcium scoring tools, which had a fixed attenuation threshold of 130 HU. Marginal voxels, which are engaged partly by stone and partly by surrounding media, are difficult to segment accurately with a fixed attenuation threshold. The ensuing volume measurement typically is either underestimated or overestimated due to threshold levels that are too high or too low, respectively. In their study using phantoms Demehri et al used variable threshold as the segmentation threshold with OsiriX software for volume estimation of urinary calculi.
机译:致编者:在讨论中,帕特尔(Patel)等人指出,在使用幻象的其他模型中,自动体积测量的准确性得到了验证。但是,他们使用的冠状动脉钙评分工具具有固定的130 HU衰减阈值。边缘体素部分地被石头和部分地被周围的介质所吸引,很难以固定的衰减阈值精确地进行分割。由于分别太高或太低的阈值水平,随后的音量测量通常被低估或高估。在他们的使用幻象的研究中,Demehri等人使用可变阈值作为OsiriX软件的分割阈值,以估算尿路结石的体积。