首页> 外文期刊>NeuroImage >Intra-and interscanner variability of magnetic resonance imaging based volumetry in multiple sclerosis
【24h】

Intra-and interscanner variability of magnetic resonance imaging based volumetry in multiple sclerosis

机译:多发性硬化症中基于磁共振成像的容积法的扫描仪内和扫描仪间变异性

获取原文
获取原文并翻译 | 示例
       

摘要

Brain volumetric measurements in multiple sclerosis (MS) reflect not only disease-specific processes but also other sources of variability. The latter has to be considered especially in multicenter and longitudinal studies. Here, we compare data generated by three different 3-Tesla magnetic resonance scanners (Philips Achieva; Siemens Verio; GE Signa MR750). We scanned two patients diagnosed with relapsing remitting MS six times per scanner within three weeks (T1w and FLAIR, 3D). We assessed T2-hyperintense lesions by an automated lesion segmentation tool and determined volumes of grey matter (GM), white matter (WM) and whole brain (GM + WM) from the lesion-filled T1-weighted images using voxel-based morphometry (SPM8/VBM8) and SIENAX (FSL). We measured cortical thickness using FreeSurfer from both, lesion-filled and original T1-weighted images. We quantified brain volume changes with SIENA.
机译:多发性硬化症(MS)中的脑容量测量不仅反映了特定疾病的过程,还反映了其他变异性来源。在多中心和纵向研究中必须特别考虑后者。在这里,我们比较了三种不同的3-Tesla磁共振扫描仪(Philips Achieva; Siemens Verio; GE Signa MR750)生成的数据。我们在三周内对每位扫描仪扫描了两次诊断为复发性MS的两名患者(T1w和FLAIR,3D)六次。我们使用自动病变分割工具评估了T2超强病变,并使用基于体素的形态计量学从病变填充的T1加权图像中确定了灰质(GM),白质(WM)和全脑(GM + WM)的体积( SPM8 / VBM8)和SIENAX(FSL)。我们使用FreeSurfer从病变填充和原始T1加权图像中测量皮质厚度。我们用SIENA量化了脑容量变化。