首页> 外文期刊>Applied optics >Charge-coupled device spectrograph for direct solar irradiance and sky radiance measurements
【24h】

Charge-coupled device spectrograph for direct solar irradiance and sky radiance measurements

机译:电荷耦合器件光谱仪,用于直接测量太阳辐射和天空辐射

获取原文
获取原文并翻译 | 示例

摘要

The characterization of a charged-coupled device (CCD) spectrograph developed at the Laboratory of Atmospheric Physics, Thessaloniki is presented. The absolute sensitivity of the instrument for direct irradiance and sky radiance measurements was determined, respectively, with an uncertainty of 4.4% and 6.6% in the UV-B, and 3% and 6% in the UV-A, visible and near-infrared (NIR) wavelength ranges. The overall uncertainty associated with the direct irradiance and the sky radiance measurements is, respectively, of the order of 5% and 7% in the UV-B, increasing to 10% for low signals [e.g., at solar zenith angles (SZAs) larger than 70%] and 4% and 6% in the UV-A, visible, and NIR. Direct solar spectral irradiance measurements from an independently calibrated spectroradiometer (Bentham DTM 300) were compared with the corresponding CCD measurements. Their agreement in the wavelength range of 310-500 nm is within 0.5% +/- 1.1% (for SZA between 20 degrees and 70 degrees). Aerosol optical depth (AOD) derived by the two instruments using direct Sun spectra and by a collocated Cimel sunphotometer [Aerosol Robotic network (AERONET)l agree to within 0.02 +/- 0.02 in the range of 315-870 mn. Significant correlation coefficients with a maximum of 0.99 in the range of 340-360 nm and a minimum of 0.90 at 870 nm were found between synchronous AOD measurements with the Bentham and the Cimel instruments. (C) 2008 Optical Society of America.
机译:介绍了由塞萨洛尼基大气物理实验室研制的电荷耦合器件(CCD)光谱仪的表征。确定了仪器用于直接辐照度和天空辐照度测量的绝对灵敏度,其中UV-B的不确定度分别为4.4%和6.6%,UV-A,可见和近红外的不确定度分别为3%和6% (NIR)波长范围。与直接辐照度和天空辐射度测量相关的总体不确定度在UV-B中分别约为5%和7%,对于低信号(例如,在太阳天顶角(SZAs)较大时,则增加到10%)大于70%],UV-A,可见光和NIR中的4%和6%。将来自独立校准的光谱仪(Bentham DTM 300)的直接太阳光谱辐照度测量结果与相应的CCD测量结果进行了比较。他们在310-500 nm波长范围内的一致性在0.5%+/- 1.1%之内(对于20度至70度之间的SZA)。两种仪器使用直接太阳光谱和并置的Cimel日光光度计[气溶胶机器人网络(AERONET)l得出的气溶胶光学深度(AOD)在315-870百万的范围内在0.02 +/- 0.02以内。在使用Bentham和Cimel仪器进行同步AOD测量之间,发现在340-360 nm范围内最大相关系数最大,在340 nm处最小相关系数为0.90。 (C)2008年美国眼镜学会。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号