首页> 外文期刊>Applied optics >High-power and high-accuracy integrating sphere radiometer: design, characterization, and calibration
【24h】

High-power and high-accuracy integrating sphere radiometer: design, characterization, and calibration

机译:大功率,高精度积分球辐射仪:设计,表征和校准

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

We describe the design, characterization, and calibration of a high-power and high-accuracy transfer standard for optical power measurements in fibers based on an integrating sphere radiometer working from -50 to +30 dBm. The integrating sphere radiometer has been calibrated in the spectral range 1250-1650 nm by use of an electrically calibrated pyroelectric radiometer and four tunable laser diodes. The total uncertainty obtained is less than +-0.8percent for these wavelength and power ranges.
机译:我们描述了基于从-50到+30 dBm工作的积分球辐射计的光纤功率测量的高功率和高精度传输标准的设计,表征和校准。积分球辐射计已通过使用电校准的热释电辐射计和四个可调激光二极管在1250至1650 nm的光谱范围内进行了校准。对于这些波长和功率范围,获得的总不确定度小于+ -0.8%。