首页> 外文期刊>Analytica chimica acta >Highly-sensitive label-free electrochemical carcinoembryonic antigen immunosensor based on a novel Au nanoparticles-graphene-chitosan nanocomposite cryogel electrode
【24h】

Highly-sensitive label-free electrochemical carcinoembryonic antigen immunosensor based on a novel Au nanoparticles-graphene-chitosan nanocomposite cryogel electrode

机译:基于新型金纳米粒子-石墨烯-壳聚糖纳米复合低温凝胶电极的高灵敏度,无标记的电化学癌胚抗原免疫传感器

获取原文
获取原文并翻译 | 示例
获取外文期刊封面目录资料

摘要

For the first time, a simple and highly sensitive label-free electrochemical carcinoembryonic antigen (CEA) immunosensor based on a cryogel electrode has been developed and tested. The as-prepared nanocomposite combined the advantages of the graphene, AuNPs and chitosan (AuNPs-GP-CS) together with the ease of preparing a cryogel coupled to a silver deposition, to act as a redox mediator, on a Au electrode. Under the optimal conditions, the decrease of the cyclic voltammetry (CV) silver peak current was proportional to the CEA concentration over a range of from 1.0 x 10(-6) to 1.0 ng mL(-1) with a detection limit of 2.0 x 10(-7) ng mL(-1). This AuNPs-GP-CS cryogel electrode gave a 1.7 times higher sensitivity and 25 times lower detection limit than the non-cryogel electrode. Moreover, the proposed electrochemical immunosensor exhibited good selectivity, reproducibility and stability. When applied to analyse clinical serum samples, the data determined by the developed immunosensor were in agreement with those obtained by the current hospital analysis system ( enzyme linked fluorescent assay) ( P > 0.05), to indicate that the immunosensor would be potentially useful for clinical diagnostics. (C) 2014 Elsevier B.V. All rights reserved.
机译:首次开发并测试了一种基于冷冻凝胶电极的简单且高度灵敏的无标记电化学癌胚抗原(CEA)免疫传感器。所制备的纳米复合材料结合了石墨烯,AuNPs和壳聚糖(AuNPs-GP-CS)的优点,以及易于制备与银沉积偶联的冷冻凝胶的功能,可在Au电极上用作氧化还原介体。在最佳条件下,循环伏安法(CV)的银峰值电流的降低与CEA浓度成正比,范围为1.0 x 10(-6)至1.0 ng mL(-1),检测极限为2.0 x 10(-7)ng mL(-1)。这种AuNPs-GP-CS低温凝胶电极的灵敏度是非低温凝胶电极的1.7倍,检测限低25倍。此外,所提出的电化学免疫传感器表现出良好的选择性,再现性和稳定性。当用于分析临床血清样品时,由开发的免疫传感器测定的数据与当前医院分析系统获得的数据一致(酶联荧光测定)(P> 0.05),表明该免疫传感器可能对临床有用诊断。 (C)2014 Elsevier B.V.保留所有权利。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号