Optimization and validation of a new method for analysis of linear alkylbenzene sulfonates in sewage sludge by liquid chromatography after microwave-assisted extraction

机译:微波辅助萃取-液相色谱法分析污泥中线性烷基苯磺酸盐的新方法的优化与验证

获取原文
获取原文并翻译 | 示例

摘要

A new approach has been developed for the monitoring of linear alkylbenzene sulfonates (LAS) (C_(10)-C_(13)) in sewage sludge.It is based on microwave-assisted extraction (MAE) prior to high performance liquid chromatography coupled with diode array (HPLC-DAD) and fluorescence (HPLC-FL) detectors.Column Zorbax XDB-C8,150 mm x 4.6 mm i.d.,5 pjn particle size was used as well as acetonitrile-water containing 0.1 M NaC1O4 (65:35) and isocratic elution.Compounds were isocratically eluted over 6min runtime at a flow rate of 1 mLmin~(-1).Since polar interferences are eluted between 0 and 2 min,they do not interfere in the analysis.The column was thermostated at a temperature of 25 °C.For the determination of LAS,DAD-UV (lambda_(ex)=225 nm) and fluorescence (lambda_(ex)=225 nm,lambda_(em)=295 run) detectors were employed.The main factors affecting the extraction efficiency (extractant volume,microwave power and irradiation time) were optimized using a compost sludge.The detection limit for total LAS in the sludge was <5 mg kg~(-1).The extraction of C_(10)-C_(13) homologues is carried out by using an extraction time of 10 min and 5 mL of methanol;whereas Soxhlet needs a 12 h extraction.The method did not require clean-up or preconcentration steps.Concentration levels of LAS were between 13037 mg kg~(-1) for digested sludge and 2492 mg kg~(-1) for compost sludge.
机译:已开发出一种新方法来监测污水污泥中的线性烷基苯磺酸盐(LAS)(C_(10)-C_(13)),该方法基于微波辅助萃取(MAE),然后进行高效液相色谱和二极管阵列(HPLC-DAD)和荧光(HPLC-FL)检测器.Zorbax XDB-C8色谱柱使用150 mm x 4.6 mm内径,5 pjn粒径以及含0.1 M NaClO4(65:35)的乙腈-水。在1minmin〜(-1)的流速下,在6min的运行时间上等度洗脱化合物。由于在0和2 min之间洗脱了极性干扰物,因此它们不会干扰分析。柱子在一定温度下恒温在25°C下测定LAS,使用DAD-UV(λ=(ex)= 225 nm)和荧光检测器(λ_(ex)= 225 nm,λ_(em)= 295 run)检测器。使用堆肥污泥优化了萃取效率(萃取剂体积,微波功率和照射时间)。污泥中的总LAS含量<5 mg kg〜(-1).C_(10)-C_(13)同源物的提取是通过10分钟的提取时间和5mL的甲醇进行的;而索氏提取需要提取12 h。该方法无需纯化或预浓缩步骤。消化污泥的LAS浓度在13037 mg kg〜(-1)至堆肥污泥的2492 mg kg〜(-1)之间。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号