首页> 外文期刊>Nuclearmedicine >Monitoring system for isolated limb perfusion based on a portable gamma camera.
【24h】

Monitoring system for isolated limb perfusion based on a portable gamma camera.

机译:基于便携式伽马相机的孤立肢体灌注监测系统。

获取原文
获取原文并翻译 | 示例
       

摘要

BACKGROUND: The treatment of malignant melanoma or sarcomas on a limb using extremity perfusion with tumour necrosis factor (TNF-alpha) and melphalan can result in a high degree of systemic toxicity if there is any leakage from the isolated blood territory of the limb into the systemic vascular territory. Leakage is currently controlled by using radiotracers and heavy external probes in a procedure that requires continuous manual calculations. The aim of this work was to develop a light, easily transportable system to monitor limb perfusion leakage by controlling systemic blood pool radioactivity with a portable gamma camera adapted for intraoperative use as an external probe, and to initiate its application in the treatment of MM patients. METHODS: A special collimator was built for maximal sensitivity. Software for acquisition and data processing in real time was developed. After testing the adequacy of the system, it was used to monitor limb perfusion leakage in 16 patients with malignant melanoma to be treated with perfusion of TNF-alpha and melphalan. RESULTS: The field of view of the detector system was 13.8 cm, which is appropriate for the monitoring, since the area to be controlled was the precordial zone. The sensitivity of the system was 257 cps/MBq. When the percentage of leakage reaches 10% the associated absolute error is +/-1%. After a mean follow-up period of 12 months, no patients have shown any significant or lasting side-effects. Partial or complete remission of lesions was seen in 9 out of 16 patients (56%) after HILP with TNF-alpha and melphalan. CONCLUSION: The detector system together with specially developed software provides a suitable automatic continuous monitoring system of any leakage that may occur during limb perfusion. This technique has been successfully implemented in patients for whom perfusion with TNF-alpha and melphalan has been indicated.
机译:背景:使用四肢灌注肿瘤坏死因子(TNF-alpha)和美法仑治疗肢体恶性黑色素瘤或肉瘤可导致高度的全身毒性,如果肢体的孤立血液区域泄漏到肢体中的话全身血管区域。当前,通过使用放射性示踪剂和笨重的外部探头来控制泄漏,该过程需要连续的手动计算。这项工作的目的是开发一种轻便,易于运输的系统,通过使用便携式伽玛相机控制全身血池放射性来监控肢体灌注泄漏,该伽玛相机适用于术中用作外部探针,并开始将其应用于MM患者的治疗。方法:建立了一个特殊的准直仪以实现最大灵敏度。开发了用于实时采集和数据处理的软件。测试该系统的适当性后,将其用于监测16例接受TNF-α和美法仑灌注治疗的恶性黑色素瘤患者的肢体灌注渗漏。结果:探测器系统的视野为13.8 cm,适合于监视,因为要控制的区域是心前区。系统的灵敏度为257 cps / MBq。当泄漏百分比达到10%时,相关的绝对误差为+/- 1%。在平均12个月的随访期后,没有患者显示出任何明显或持久的副作用。接受TNF-α和美法仑的HILP后16例患者中有9例(56%)观察到部分或完全病变。结论:该检测器系统与专门开发的软件一起提供了一种合适的自动连续监测系统,可对肢体灌注期间可能发生的任何泄漏进行监测。该技术已成功应用于已灌注TNF-α和美法仑的患者。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号